Graseby 3400
The Graseby 3400 Syringe Pump is the gold standard pump for pharmacological stress testing. It can infuse virtually any IV drug within a range of 0.1-1200 ml/hr, has multiple unit selection, syringe size flexibility, is AC or battery operated and has an accuracy of ±2%. Typical IV drugs infused with the 3400 are Adenosine, Dobutamine, Persantine, Zevalin, Cholecystokinan (Kinevac) and more.
Features
- ONE PUMP for infusion of Adenosine, Dobutomine, Persantine, Zevalin, Bexxar and Kinevac
- Wide infusion rate range: 0.1-1200 ml/hr
- Multiple unit selection: ml/hr, µg/hr, mg/hr, µg/kg/hr, mg/kg/hr, µg/kg/mn, mg/kg/mn
- Syringe size flexibility: 5ml, 10ml, 20ml, 30ml, 50/60ml
- Simple operation: Menu driven display
- AC or battery operated (internal rechargeable battery)
Call for price:
(855) 477-8866
Description
KoDimensions: 325 mm (long) x 195 mm (high) x 115 mm (deep) with the pole clamp fitted and the plunger clamp closed.
Weight: 3.5 kg including batteries and pole clamp.
AC supply: 100 to 240 V, 50/60 Hz, 40 VA. The power supply uses Primary Switching in order to utilise the AC supplies of most countries.
Battery type: Sealed lead acid, rechargeable (Cyclon, 3 off). SIMS Graseby recommend that the batteries are checked at least annually.
Battery life: More than 10 hours of normal pump operation when the batteries are fully charged. With the AC supply connected, up to 14 hours will be required to fully recharge low voltage batteries.
Syringe brands:
BD Plastipak 5;10; 20; 30/35 or 50/60 ml.
Terumo 5; 10; 20; 30/35 or 50/60 ml.
Braun Omnifix 5;10; 20; 30/35 or 50/60 ml.
Sherwood Monoject 5; 10; 20; 30/35 or 50/60 ml.
IMS Pump-jet 30 30 ml (prefilled).
Fresenius Injectomat 50 ml.
Braun Perfusor 50 ml
Flow rate:
Continuous; 0.1 to 400 ml/h in 0.1 ml increments.
Mass units; mg/kg/h etc.
Bolus; 0.1 to 1200 ml/h in 0.1 ml increments.
Totaliser: 0 to 999.9 ml in 0.1 ml increments.
Adjustable occlusion pressure: 1.85 kg (250 mm Hg) to 7.42 kg (1000 mm Hg).
Operating conditions:
+5° to +40°C, 30 to 75% Rh, 700 to 1060 hPa.
Storage conditions:
-40° to +70°C, 30 to 90% Rh, 700 to 1060 hPa.
Drive accuracy: ± 2% when measured over a complete syringe.
Design standards: IEC 601-1.
Elec. safety: Class II; Type CF.
Languages available: Selectable via the configuration mode.