Abbott Lifecare PCA 3
The LifeCare PCA 3 Infusion System is the most advanced PCA device to help address wrong dose and wrong drug errors. The LifeCare PCA 3 uses an integrated approach to patient-controlled analgesia and helps to enhance patient safety and efficiency through bar code technology and advanced software design. Pre-Filled bar-coded drug syringes provide quality and enhanced safety.
Features
- Designed to help reduce medication errors, enhance utility and increase efficiency.
- Built-in bar code reader automatically identifies drug name and drug concentration.
- Review screens confirm data input before infusion begins.
- Advanced software, soft-function keys and numeric keypad Stored speed protocol function saves time and increases compliance with standing orders.
- Designed to help reduce medication errors, enhance utility and increase efficiency.
- Built-in bar code reader automatically identifies drug name and drug concentration.
- Review screens confirm data input before infusion begins.
- Advanced software, soft-function keys and numeric keypad Stored speed protocol function saves time and increases compliance with standing orders.
Call for price:
(855) 477-8866
Description
PHYSICAL
- Dimensions: Approximately 8 W x 13 H x 6 D in. (excluding pole clamp, power cord)
- Weight: Approximately 10 lbs. with battery
- Casing: High-impact plastic
ELECTRICAL
- Power Requirements: 105-130 V, 57 – 63 Hz, 50 W
- Power Cord: Hospital-grade AC cord. 10 feet with transparent plug and retainer plate
- Fuses: 0.5 A, 250 V, Slo-Blo
- Battery: One sealed lead acid, rechargeable 8 V battery, internal to the device
- Battery Operation: A fully charged new battery provides four hours of operation and delivers 30 mL of solution. Operation time is measured from initial pumping to the LOW BATTERY alarm.
- Recharge: The battery charges whenever the infusion pump is connected to AC power. After a LOW BATTERY alarm, the battery will recharge to 80% of full capacity within 16 hours.
- Self-Discharge: A fully charged battery will retain a minimum 20% of the initial charge for six months when the infusion pump is not connected to AC power or is not operating, and is stored at 20° C.
ENVIRONMENT
- Operating: 41° to 104° F (5° to 40° C) 10% to 90% relative humidity
- Transporting and Storage: -4° to 140° F (-20° to 60° C) 10% to 90% relative humidity
- Atmospheric Pressure: 0 – 10,000 feet (0 – 3000 meters) or equivalent atmospheric pressure
- Relative Humidity: 10 – 90% non-condensing up to 104° F (40° C); maximum of 15% non-condensing above 104° F (40° C)
OCCLUSION ALARM AND LIMITS
- Backpressure Range: 10 to 20 psi
- MODES
- PCA Only: Approximately 1 mL in 35 seconds
- Continuous Only: Variable from 0.1 – 20x concentration (mg/hr or mcg/hr)
- PCA + Continuous: Variable from 0.1 – 20x concentration (mg/hr or mcg/hr)
- Lockout Interval Range: 5 to 120 minutes in one minute increments