Curlin 4000 CMS
The Moog Curlin 4000 CMS infusion pump is an easy-to-use, ambulatory or pole-mounted infusion device designed to meet the needs of the hospital or alternate site patient. It operates accurately in any position, has a low rate resolution of 2 microliters, and its small, compact, lightweight design allows mobility for ambulatory patients. The Curlin Medical pump can be carried in any of the four convenient carryall soft-packs, pole mounted, placed into the locked “safety shells,” or set on a suitable surface for use.
Features
- Cost effective, safe and easy-to-load disposable Curlin Medical administration sets featuring a unique, spring-activated, self-clamping, set-based “Integral Flow-Stop” device that automatically clamps the tubing when the door of the pump is opened and prevents inadvertent fluid flow.
- User-friendly, easy-to-teach programming that shortens staff in service time and patient teaching time.
- Three tamper-resistant lock settings to assist in maintaining patient compliance and safety.
- Selection by user from three units of delivery, milliliters (ml), milligrams (mg), and micrograms (mcg).
- Retention of programmed infusion settings until cleared by the clinician.
- Mandatory requirement of two independent key press actions before a running pump can be turned off, thus minimizing the accidental interruption of a therapy in progress.
- Audio alarms that can be adjusted from 1 (very quiet) to 9 (loudest) to meet specific patient and clinical setting needs, but that cannot be totally disabled.
- Retained memory features that allow the pump to easily resume a therapy exactly from where it left off when it is interrupted before completion.
- Pump will not permit the user to enter any programming value that is outside of its predetermined range.
- Integral High Upstream pressure alarm to notify user of excessive upstream pressure.
Call for price:
(855) 477-8866
Description
- Pumping Mechanism: Curvilinear peristaltic finger action
- Dimensions: 5 H x 4 W x 2.5 D in.
- Weight: 17.5 oz
- Case Material: UL rated flame-retardant, non-toxic high impact plastic
Operating Temperature:
- ± 6% Accuracy: 55° to 90°F
- ±10% Accuracy: 45° to 55°F
- ±10% Accuracy: 90° to 104°F
Operating Relative Humidity: 15 to 95% non-condensing
Storage Temperature Range: -20° to 140°F
Storage Relative Humidity: 5 to 95% non-condensing
Power Requirements: 2 “C size” Single-Use Batteries or AC Power
Operating Time with Full Charge Batteries:
- Nominal 10 hours at a rate of 400 ml/hr
- Nominal 30 hours at a rate of 125 ml/hr
- Nominal 85 hours at a rate of 2 ml/hr
Internal Memory Battery: 5 years life
Display: Graphic LCD, 4 text fields, 16 characters per field
Flow Rates: 0.1 to 400 ml/hr
KVO Rates: 0 to 10 ml/hr. Can be set at “0” in Continuous, TPN, Variable and Intermittent Therapies
Volume Accuracy:
± 6% for volume delivery >1 ml under the following conditions:
- Temperature 55° to 90° F
- Back pressure to 300 mmHg (5.8 PSI)
±10% for volume delivery >1 ml under the following conditions:
- Temperature 45° to 55°F
- Temperature 90° to 104°F
- Back pressure to 750 mmHg (14.5 PSI)
Recommended administration set usage not to exceed 72* hours.
Administration Sets: Proprietary Curlin Medical non-DEHP, PVC administration sets available in different configurations.
Fill Volume for Administration Sets: Nonfiltered microbore sets ± 4 ml. See individual administration set label for specific fill volume
Up Occlusion Detection Time:
Microbore tubing with occlusion below spike
- <20 minutes at 0.1 ml/hr
- < 4 minutes at 1 ml/hr
- <20 seconds at 10 ml/hr
- <10 seconds at 50 ml/hr