CADD Solis
The CADD-Solis® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-Solis® pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia).
Features
- Small and lightweight pump for convenience
- Programmed intermittent bolus
- Innovative therapies and task-based menus
- Scalable platform
- User-friendly interface
- Graphical reports
- Safety software
Call for price:
(855) 477-8866
Description
- Dimensions: 1.6″ W X 4″ H X 5″ L
- Weight: 1.31 lbs. (21 oz), including 4 AA batteries
- Resolution: Medication Cassette Reservoir or CADD™ Administration Set, 0.050 ml per pump stroke nominal
- Moisture Protection: Splash-proof (IPX4) per IEC 60529
- Power: 4 AA alkaline batteries; AC adapter; rechargeable battery pack
- Internal Memory Backup: It charges whenever the pump is powered on and has a 10 month memory capacity once it has been charged for 250 hours at 68°F
- System Operating Temperature: 36°F to 104°F (2°C to 40°C)
- System Storage Temperature: -4°F to 140°F (-20°C to 60°C)
- Atmospheric Pressure: 70 kPa (10.2 psi) to 106 kPa (15.4 psi)
- Maximum Infusion Pressure: 27.0 psi
- Maximum Time to Occlusion Alarm: 0.1 ml/hr – 122 min; 10 ml/hr – 122 sec
- Bolus Volume at Occlusion Alarm Pressure:
- 0.107 ml reservoir at 0.01 ml/hr: <0.136 ml
- 0.116 ml reservoir at 10 ml/hr: <0.124 ml
- Accuracy: ± 6% (nominal)
- High Pressure Alarm Threshold: 18 ± 9 psi
- Air Detector Alarm:
Sensitivity:
-
- Low – Single bubble greater than 400 uL
- High – Single bubble greater than 150 uL
Accumulated Air – Greater than 1 ml air over 15 minutes (nominal)
- Maximum Volume Infused Under Single-Fault Conditions: CADD® administration set: 0.15 ml
- Delivery Rate during priming: Approx. 250 ml/hr
- Alarm disabled during priming: Air-in-Line
Delivery Specifications:
- Reservoir Volume: 0 to 9999; programmable in 1 ml increments, displayed in 0.1 ml increments
- Units: milliliters (ml), milligrams (mg), micrograms (mcg)
- Concentration:
- Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, …15, 20, 25, …100
- Mcg/ml: 1, 2, 3, …15, 20, 25, …100, 200, 300, 400, 500
- (Default = 1)
- Continuous Rate: 0-30 ml/hr (or the mg or mcg equivalent)
- PCA Dose:
- 0 to 20 ml (or the mg or mcg equivalent)
- Delivery rate (Continuous Rate + PCA Dose): programmable from 40-175 ml/hr
- PCA Dose Lockout: 1 minute to 24 hours in the following increments:
- 1 minute for values between 1 and 20 minutes
- 5 minutes for values between 20 minutes and 24 hours
- Max Doses per Hour: 1 – 60 doses
- Delivery Limit Amount:1 to 1000 ml (or mg or mcg equivalent) in increments of:
- 01 ml from 0.01-0.5 ml
- 5 ml from 0.5-100 ml
- 0 ml from 100-1000 ml
- Given: 0 to 99,999.99 in 0.01 unit increments
- Clinician Bolus:
- 0 – 20 ml (or mg or mcg equivalent)
- Delivery rate (Continuous Rate + Clinician Bolus): programmable from 40-175 ml/hr
Alarms/Alerts:
Battery depleted while delivering | Reservoir volume empty |
Battery removed while delivering | Pressure sensor faulty |
Battery unusable while delivering | Downstream occlusion |
Rechargeable battery end of life | Upstream occlusion |
Download the operator’s Manual