CADD Legacy
The CADD-Legacy™ 1 ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-Legacy™ 1 pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion.
Features
- Easy to read display
- Cassette detection
- Occlusion detection
- Air-in-Line detector with Off, High- and Low-sensitivity settings
- ON/OFF key
- Event memory
- Reliability
- Programming flexibility
Call for price:
(855) 477-8866
Description
- Dimensions: 1.6″ W X 3.8″ H X 4.4″ L
- Weight: 0.86 lbs. (13.8 oz), including 2 AA batteries and empty medication reservoir
- Resolution: Medication Cassette Reservoir or CADD™ Administration Set, 0.050 ml per pump stroke nominal
- Moisture Protection: Splash-proof (IPX4)
- Power: Two AA alkaline batteries; AC Adapter
- Battery Capacity: The expected life of 2 AA batteries is 15 hours at 125 ml/hour, or approximately 14 days at 10 ml/day (nominal). It is recommended that 2 new AA batteries be kept available for replacement if necessary.
- Accuracy: ± 6% (nominal). At low infusion rates, this accuracy may not be achieved for short periods.
- Maximum Infusion Pressure: 40.0 psi
- Maximum Time to Occlusion Alarm: CADD® Administration Set with anti-siphon valve: 2.0 hours
- Bolus Volume at Occlusion Alarm Pressure: 0.050 ml resolution CADD® Administration Sets/Medication Cassette™ Reservoirs: <0.15 ml
- Operating Temperature: +35°F to 104°F (2°C to 40°C)
- Storage/Transportation Temperature: -4°F to 140°F (-20°C to 60°C)
- High Pressure Alarm: 26 ± 14 psi
- Air Detector Alarm:
- Single bubble Low Sensitivity: > 0.250 ml
- Single bubble High Sensitivity: > 0.100 ml
- Multi-bubble: 1.0 ml nominal
- Maximum Volume Infused under Single Fault Condition: CADD® Administration Set with anti-siphon valve: 0.2 ml
- Delivery Rate during priming: Approximately 180 ml/hr
- Alarm Disabled during priming: Air Detector
- Continuous Delivery Specifications:
- Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
- Continuous Rate: 1 to 3000 ml/24 hr; programmable in 1 ml/24 hr increments (Default = 1.0 ml)
- 0 to 99999.95 in 0.05 ml increments
Alarms/Alerts
- Low battery power
- Depleted battery power
- Battery dislodged
- Pump stopped
- Pump fault
- Low reservoir volume
- High delivery pressure
- Air in line
- Motor locked
- Upstream occlusion
- Reservoir volume empty
- Key stuck
- Disposable detached