Baxter Sigma Spectrum Version 6
The Sigma Spectrum is intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products, and mixtures of required patient therapy. The intended routes of administration consist of the following clinically acceptable routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. The Spectrum is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by the user.
Brand new units may be available. Please contact us directly for more information.
- Flow Rate Range: 0.5 to 999 mL/hr with 0.1 mL/hr increments
- Flow Rate Accuracy: 0.5 – 1.9 mL/hr: ±0.1 mL/hr; 2.0 – 800 mL/hr: ±5%; 801 – 999 mL/hr: ±5%
- Flow Rate Tolerance: ± 5%Delivery Volume: 0.01 – 9,999 mL
- Occlusion Pressure: (Adjustable) High (19 ±9 PSI), Medium (13 ±6 PSI), and Low (6 ±4 PSI)
- KVOs: 0.5 – 50 mL/hr
- Power Requirement: Lithium Ion – 1800 mA/h, 7.4 VDC nominal
- Battery Operating Time: 4 Hrs (Charge Time: 16 Hrs)
Software Versions:
- 6.05
- 6.02
- Dimensions: 6.3 in. L x 4.2 in. W x 2.5 in. D (16 cm. L x 10.67 cm. W x 6.35 cm. D)
- Weight: 2.2 lbs (0.5 kg) ± 0.06 lbs for Accessories
- Parts Included: AC Adapter, Battery, Pole Clamp
- Display Resolution: Color, 240 x 270, LED Front-Lit (0.2235 mm X 0.2235 mm)
More Details:
- Pumping Mechanism: Linear peristaltic
- Alarm Volume: 3 levels – high, medium, and low
- Alarms & Alerts: Air-in-line, Downstream Occlusion, Very Low Battery, Due for inspection
- Logging Memory: 10,000 ± event capacity; Separate pump history log and drug event log
- Anti-Free-Flow System: Set-based, utilizing IV set slide clamp.
| Weight | 3 lbs |
|---|---|
| Dimensions | 12 × 12 × 12 in |
| Product Condition |
- Pole Clamp (New or Used)
- AC Adapter
- B Wireless Battery