Curlin Painsmart IOD
The Moog Curlin PainSmart IOD (Information On Demand) is a compact ambulatory infusion pump built on the proven Curlin 6000 platform and designed for PCA (Patient Controlled Analgesia), PCEA, and subcutaneous pain management therapies. The IOD model adds dedicated bedside buttons that allow clinicians and patients to view real-time infusion data — hourly totals, boluses attempted and given, shift totals, and line pressure — without pausing the running therapy. The PainSmart IOD supports intravenous, epidural, and subcutaneous administration routes with delivery programmable in milliliters, milligrams, or micrograms.
Flow rates range from 0.1 to 400 ml/h with ±5% accuracy, and built-in medication limits (medLMTS™) prevent programming values outside of preset safe ranges. When paired with Moog’s Protocol Library Safety System (PLSS), clinicians can upload pre-validated drug protocols to further reduce programming errors. The pump runs on two standard C-cell alkaline batteries with operating times up to 85 hours at 2 ml/h, and accepts an optional rechargeable Li-Ion battery pack and AC adapter for extended or facility-based use.
IPR Medical buys, sells, repairs, and stocks parts for the Moog Curlin PainSmart IOD. Whether you need a recertified pump, replacement parts, or a repair on an existing unit, contact us at 855-477-8866 or cs@iprmed.com.
Moog Curlin PainSmart™ IOD — Ambulatory PCA Infusion Pump with Info-On-Demand
Compact ambulatory PCA infusion pump with dedicated Info-On-Demand (IOD) buttons for real-time bedside data access without pausing the infusion. Built on the Curlin 6000 platform with curvilinear peristaltic mechanism and 2-microliter flow resolution. Supports IV, epidural, and subcutaneous routes. 0.1–400 ml/h flow rate, ±5% accuracy, built-in medication limits (medLMTS™), and PLSS protocol library compatibility. 1.1 lbs, IPX1 water-resistant, powered by two standard C-cell batteries with up to 85 hours of operating time.
| Pump Platform | Moog Curlin 6000 Series |
| Model Variant | PainSmart IOD (Information On Demand) |
| Intended Use | PCA (Patient Controlled Analgesia), PCEA, and subcutaneous pain management infusion — hospital, alternate-site, and ambulatory settings |
| IOD Feature | Dedicated Info-On-Demand buttons provide real-time access to hourly totals, boluses attempted/given, shift totals, and line pressure without pausing the infusion |
| Pumping Mechanism | Patented curvilinear peristaltic; 2-microliter flow resolution |
| Administration Routes | Intravenous (IV), Epidural (EPI), Subcutaneous (SQ) — routes can be institution-fixed via software |
| Units of Delivery | ml, mg, mcg — units can be institution-fixed via software |
| Flow Rate Range | 0.1–400 ml/h (0.1 ml/h increments) |
| PCA Basal Rate Limits | IV: 0.1–50 ml/h max | EPI: 0.1–25 ml/h max | SQ: 0.1–5 ml/h max |
| Accuracy | ±5% (with proprietary Curlin administration sets) |
| Volume Limits (VTBI) | 1.0–9,999 ml (0.1 ml increments) |
| KVO Rate | 0.0–10 ml/h (not available in PCA mode) |
| Patient Bolus Dose | 0.1–9.9 ml/dose |
| Maximum Boluses per Hour | Up to 15 |
| Bolus Lockout Interval | 1–60 minutes (1-minute increments) |
| Single Fault Bolus Volume | ≤0.8 ml |
| Medication Limits | medLMTS™ — customizable maximum basal rate, bolus, clinician dose, and load dose limits |
| Protocol Library | PLSS (Protocol Library Safety System) — upload pre-validated drug protocols; 100+ protocol capacity |
| Display | 2.5″ × 1.2″ LCD with Tallman Lettering; continuous backlight on AC power |
| Security / Lock Levels | 3 tamper-resistant lock settings; 5 or 9 digit clinician access code; unused therapies can be blocked |
| Data Log | 6,000 events — includes line pressure, volume infused, time stamps; downloadable via Patient History Manager (PHM) or PLSS software |
| Occlusion Detection | Upstream and downstream; adjustable downstream settings (High / Low) |
| Alarms | Air-in-line, high upstream pressure, occlusion upstream, occlusion downstream, door open, set not installed, unattended pump, replace set, empty battery, infusion complete |
| Alarm Volume | Adjustable (levels 1–9); cannot be fully disabled |
| Administration Sets | Proprietary Curlin non-DEHP PVC sets; latex-free; integral flow-stop free-flow protection; configurations include microbore, filtered (0.22 and 1.2 micron), yellow-striped epidural sets, and sets with 150 ml integral reservoirs |
| Admin Set Max Usage | 72 hours (per standards of practice) |
| Water Resistance | IPX1 (protected against dripping water); fluid-resistant housing — not submersible |
| Electrical Classification | Defibrillation-proof type CF applied parts; continuous mode of operation |
| Power — Primary | 2 × C-cell alkaline batteries (standard) |
| Battery Operating Time | ~10 hrs at 400 ml/h | ~30 hrs at 125 ml/h | ~85 hrs at 2 ml/h |
| Power — External (Optional) | Rechargeable Li-Ion battery pack (360-2030); Curlin 6000 AC adapter/charger |
| Dimensions | 5.1″ × 4.0″ × 2.5″ (13 × 10.2 × 6.4 cm) |
| Weight | 1.1 lbs (18.1 oz / 513 g) |
| MRI Safety | NOT MRI safe — do not enter the MRI scanner room (projectile hazard) |
| Safety Standards | UL 60601-1; CSA C22.2 No. 601-1; EN 60601-1; EN 60601-1-2; EN 60601-2-24 |
| Accessories | Detachable pole clamp, remote bolus cord, lockable safety shells (multiple sizes), pump holster, keypad covers, carry packs (4 sizes) |
| Maintenance | Annual preventive maintenance recommended; maintenance due date auto-set after 12 hours of initial use |
| Manufacturer | Moog Medical Devices Group (manufactured by ZEVEX, Inc., Salt Lake City, UT) |
IPR Medical buys, sells, repairs, and stocks parts for the Moog Curlin PainSmart IOD. Contact us at 855-477-8866 or cs@iprmed.com.
Trademark Notice: Moog, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Moog.
Info-On-Demand — real-time bedside data without pausing the infusion
The PainSmart IOD’s defining feature is its dedicated Info-On-Demand buttons. While the pump is running, clinicians or patients can press an IOD button to instantly view hourly totals, boluses attempted and given, shift totals, and downstream line pressure — all without pausing the infusion. The requested data displays for several seconds and the screen returns to the standard run view automatically. This gives clinicians actionable pain management data at the bedside and supports faster titration decisions, better documentation during rounds, and more informed patient communication. The IOD feature is not available on the standard PainSmart model.
Purpose-built for PCA, PCEA, and subcutaneous pain therapy
The PainSmart IOD is an ambulatory infusion pump designed specifically for patient-controlled analgesia (PCA), patient-controlled epidural analgesia (PCEA), and subcutaneous pain management. Built on the proven Curlin 6000 platform, it supports three administration routes — intravenous, epidural, and subcutaneous — with delivery programmable in milliliters, milligrams, or micrograms. Institutions can lock the pump to a specific route and unit of measure via software configuration, ensuring bedside staff program within approved parameters. At 1.1 lbs and 5.1 × 4.0 × 2.5 inches, the PainSmart IOD is compact enough to be carried in one of four available soft carry packs, pole-mounted, holstered, or placed inside a lockable safety shell.
Curvilinear peristaltic mechanism — smoother delivery at every rate
The PainSmart IOD uses Moog’s patented curvilinear peristaltic pumping mechanism, which delivers fluid in 2-microliter increments — significantly finer than conventional linear peristaltic pumps. This translates to smoother, more consistent flow at all rates, with particular advantages at the low rates common in PCA and epidural infusions where continuity of flow directly impacts patient comfort and drug efficacy. Flow rates range from 0.1 to 400 ml/h in 0.1 ml/h increments with ±5% accuracy when used with proprietary Curlin administration sets. PCA basal rate maximums are route-specific: up to 50 ml/h IV, 25 ml/h epidural, and 5 ml/h subcutaneous.
Built-in medication limits and protocol library support
The PainSmart IOD features medLMTS™ — customizable medication limits that prevent programming values outside of predetermined safe ranges for basal rate, bolus dose, clinician dose, and load dose. Patient bolus doses are programmable from 0.1 to 9.9 ml with a lockout interval of 1 to 60 minutes and a maximum of 15 boluses per hour. When paired with Moog’s Protocol Library Safety System (PLSS), the pump accepts pre-validated drug protocols uploaded via software — eliminating manual programming for common regimens and reducing the opportunity for dosing errors. PLSS supports over 100 protocols organized by clinical category, and PCA dose schedules and medLMTS™ limits can be maintained across therapy repeats.
Security, safety, and comprehensive alarm system
Three tamper-resistant lock levels — controlled by a 5 or 9 digit clinician access code — allow facilities to restrict patient access to programming functions while maintaining clinician control. Unused therapies can be blocked entirely from the programming menu. The PainSmart IOD’s integral alarm system monitors for air-in-line, upstream and downstream occlusion, high upstream pressure, door open, set not installed, unattended pump, replace set, empty battery, and infusion complete conditions. Downstream occlusion sensitivity is adjustable between High and Low settings. The maximum volume that may be infused under single fault conditions is 0.8 ml. Alarm volume is adjustable across nine levels but cannot be fully disabled. A 6,000-event data log records line pressure, volume infused, and time stamps — downloadable via Patient History Manager (PHM) or PLSS software for compliance documentation and clinical review.
Up to 85 hours on two standard C-cell batteries
The PainSmart IOD runs on two readily available C-cell alkaline batteries — no proprietary battery packs required for standard operation. Operating times range from approximately 10 hours at 400 ml/h to 85 hours at 2 ml/h, making multi-day ambulatory therapy practical on a single set of batteries. For facility-based or extended use, an optional rechargeable lithium-ion battery pack and AC adapter/charger are available. The pump’s large LCD display features Tallman Lettering for improved drug name readability, with continuous backlighting when connected to AC power. A mandatory two-key-press sequence to power off during a running infusion prevents accidental therapy interruption. IPX1 water resistance protects against dripping water, and the impact-resistant housing is designed for the demands of ambulatory patient use.
Proprietary administration sets with built-in free-flow protection
The PainSmart IOD uses proprietary Curlin administration sets manufactured from non-DEHP PVC without natural rubber latex. Every set features Moog’s integral flow-stop device — a spring-activated, self-clamping mechanism that automatically occludes the tubing when the pump door is opened, preventing inadvertent free flow. Set configurations include standard microbore, 0.22 and 1.2 micron filtered sets, yellow-striped epidural-specific sets, and sets with 150 ml integral fluid reservoirs. Administration set usage should not exceed 72 hours per standards of practice. Anti-siphon valve (ASV) sets are available for additional free-flow protection in clinical situations where it is indicated.
IPR Medical — sales, repairs, and parts
IPR Medical buys, sells, repairs, and stocks parts for the Moog Curlin PainSmart IOD. Whether you need a recertified pump ready for immediate deployment, replacement parts to keep your existing fleet running, or a full repair and recertification, IPR Medical’s infusion pump service team has the expertise and inventory to support your program. Annual preventive maintenance is recommended by the manufacturer — contact IPR Medical to schedule service or request a quote.
IPR Medical buys, sells, repairs, and stocks parts for the Moog Curlin PainSmart IOD. Contact us at 855-477-8866 or cs@iprmed.com for pricing, repairs, parts, or technical support.
Trademark Notice: Moog, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Moog.
| Weight | 0.9 lbs |
|---|---|
| Dimensions | 2.5 × 4 × 5 in |
| Condition | Recertified (Refurbished) |
- AC Power Adapter
- Bolus/ Patient Cable
- Lock Box (500 mL)
- Pole Clamp
- Holster
Reviews
There are no reviews yet.