Curlin 6000 CMS
The Moog Curlin 6000 CMS is a compact, ambulatory infusion pump built for both hospital and home infusion environments. Its patented curvilinear peristaltic design delivers fluid in smaller, more precise increments than standard linear peristaltic pumps — an advantage that matters most when managing sensitive patient populations, short half-life medications, or low-rate infusions. With five therapy modes — Continuous, PCA, TPN, Intermittent, and Variable — the Curlin 6000 handles the full range of clinical infusion scenarios from a single, lightweight device that weighs just 18.1 oz. At ±5% accuracy and a 2 mcL flow resolution, it sets a high bar for ambulatory-class performance.
The pump supports dosing in mL, mg, and mcg, and integrates with Moog’s Protocol Library Safety System (PLSS) to streamline programming and reduce the risk of user error. It is powered by two C-cell alkaline batteries and can also run on AC power or an external rechargeable battery pack for extended use. The Curlin 6000 requires Moog Curlin proprietary non-DEHP, latex-free administration sets — IPR Medical stocks these sets, so customers can source both pump and consumables in one place.
IPR Medical buys, sells, repairs, and stocks parts for the Moog Curlin 6000 CMS. Whether you need a fully serviced unit, a repair on a pump already in your fleet, or replacement administration sets, our team is ready to help. Call 855-477-8866 or email cs@iprmed.com.
Moog Curlin 6000 CMS — Technical Specifications
| Pump Type | Ambulatory volumetric infusion pump |
| Pumping Mechanism | Patented curvilinear peristaltic finger action |
| Therapy Modes | Continuous, PCA (Patient-Controlled Analgesia), TPN with Automatic Ramping, Intermittent, Variable (up to 24 programs) |
| Administration Routes | Intravenous (IV), epidural (EPI), subcutaneous (SQ), intra-arterial, enteral |
| Flow Rate Range | 0.1 – 400 ml/h (IV); increments of 0.1 ml/h PCA: 0.1 – 50 ml/h IV; 0.1 – 25 ml/h EPI; 0.1 – 5 ml/h SQ Note: PLSS required for IV rates above 9.9 ml/h |
| Flow Rate Resolution | 2 mcL (microliters) |
| Accuracy | ±5% (for volume delivery >1 ml); accuracy may vary when used with syringes |
| Volume Limits | 1.0 – 9,999 ml; increments of 0.1 ml |
| KVO Rate | 0.0 – 10 ml/h (not available in PCA mode) |
| Dosing Units | mL, mg, mcg |
| PCA Bolus Volume | 0.1 – 9.9 ml/dose (IV); 0.1 – 25 ml/dose (EPI); 0.1 – 5 ml/dose (SQ) |
| Max Boluses | 15 per hour |
| Bolus Lockout | 1 – 60 minutes; increments of 1 minute |
| Display | 2.5″ × 1.2″ backlit LCD; continuous backlight when on AC power; Tallman lettering |
| Alarms | Upstream occlusion, downstream occlusion, air-in-line, high upstream pressure, door open, empty battery, infusion complete; adjustable audio volume (1–9) |
| Power Sources | 2 × C-cell alkaline batteries (included); AC adapter/charger; optional external rechargeable battery pack |
| Water/Fluid Resistance | IPX1 (protected against dripping water); impact-resistant housing; not for submersion |
| Operating Environment | Relative humidity: 15–95% non-condensing |
| Dimensions | 5.1″ H × 4.0″ W × 2.5″ D (13 cm × 10.2 cm × 6.4 cm) |
| Weight | 18.1 oz (513 g) |
| Safety Features | Integral Flow-Stop (auto-clamps on door open), MedLMTS™ dose limits, three tamper-resistant lock settings, two-keypress pump-off confirmation, memory retention on power interruption |
| Drug Library / PLSS | Compatible with Protocol Library Safety System (PLSS) v2 for pre-validated drug protocols and patient-specific prescriptions |
| Administration Sets | Proprietary Moog Curlin non-DEHP, latex-free PVC sets with integral Integral Flow-Stop; max recommended use 72 hours. IPR Medical stocks Moog Curlin administration sets. |
| Regulatory | UL 60601-1; CSA C22.2 No. 601-1; EN 60601-2-24; Type CF defibrillation-proof applied parts; IPX1 |
| FDA Device Corrections | Class I recall (2011) for EC45 software error on units manufactured/updated May 2007–Feb 2011. See Documents tab and contact IPR Medical for service history and software status. |
IPR Medical buys, sells, repairs, and stocks parts and IV sets for the Moog Curlin 6000 CMS. Contact us at 855-477-8866 or cs@iprmed.com.
About the Moog Curlin 6000 CMS
The Moog Curlin 6000 CMS is a true multi-therapy ambulatory infusion pump, and for nearly two decades it has been a trusted workhorse in home infusion, hospital, and long-term care settings. Its compact dimensions — just 5.1″ × 4.0″ × 2.5″ and 18.1 oz — make it easy to carry in a soft pack or mount on a pole without restricting patient mobility. Behind that small footprint is a sophisticated delivery engine: a patented curvilinear peristaltic mechanism that produces 25 times finer flow resolution than standard linear peristaltic pumps. That difference in Resolution of Flow and Continuity of Flow directly affects how consistently medication reaches the patient, especially during low-rate infusions or when managing drugs with short half-lives.
Five Therapy Modes for Maximum Clinical Flexibility
The Curlin 6000 CMS supports five therapy delivery modes from a single platform. Continuous therapy delivers a steady, programmed rate without interruption. PCA (Patient-Controlled Analgesia) mode allows clinician-configured basal rates with patient-demand boluses on a lockout schedule — supporting IV, epidural, and subcutaneous routes. TPN with Automatic Ramping manages parenteral nutrition with optional taper-up and taper-down periods to reduce metabolic stress. Intermittent mode delivers medication in scheduled intervals at specified rates, with an optional keep-vein-open (KVO) rate between doses. Variable therapy allows up to 24 distinct programs of varying rates, volumes, and durations within a single prescription. This range of modes makes the 6000 CMS suitable for oncology, pain management, antibiotic therapy, nutritional support, and a broad spectrum of outpatient and home infusion applications.
Safety Architecture and Programming Controls
Safety on the Curlin 6000 CMS is built in at multiple layers. The Integral Flow-Stop automatically clamps the tubing the moment the pump door is opened, preventing free-flow events. The MedLMTS™ system allows institutions to set customized maximum dose, basal rate, and bolus limits at the BioMed configuration level, so clinicians work within pre-established guardrails. Three tamper-resistant lock settings protect programming from unintended patient or caregiver changes. The pump also requires two independent keypress actions to power off during a running infusion, reducing the chance of accidental therapy interruption. Programmed infusion settings are retained in memory until the clinician actively clears them, eliminating the need to re-enter a prescription unnecessarily. Alarms cover upstream and downstream occlusion, air-in-line detection, high upstream pressure, door-open status, low battery, and infusion complete — with an adjustable audio volume from level 1 to 9 to match the clinical environment.
Protocol Library Safety System (PLSS)
For facilities managing complex or high-risk drug regimens, the optional Protocol Library Safety System (PLSS) adds another layer of protection. PLSS allows administrators and pharmacists to pre-load validated drug protocols or patient-specific prescriptions directly onto the pump, eliminating manual rate and concentration entry at the bedside. This shortens programming time, reduces the opportunity for user error, and is required to enable IV flow rates above 9.9 ml/h on this platform. PLSS also supports patient history logging and report generation for compliance and quality review purposes. IPR Medical can assist with PLSS-related service and configuration questions.
IV Sets and Consumables
The Curlin 6000 CMS requires Moog Curlin proprietary administration sets — non-DEHP, latex-free PVC sets with an integrated Integral Flow-Stop device. These sets are not interchangeable with generic tubing; use of non-proprietary sets can alter delivery accuracy and create patient safety risk. Sets are available in multiple configurations and are recommended for replacement every 72 hours. IPR Medical stocks Moog Curlin administration sets, making it easy for facilities to source both pump service and consumables from a single vendor. Consolidating your pump and consumable supply through IPR reduces procurement overhead and ensures you always have compatible sets on hand.
FDA Device Corrections and Safety Bulletins
In 2011, the FDA classified a voluntary Moog correction of the Curlin 6000 CMS, 6000 CMS IOD, PainSmart, and PainSmart IOD as a Class I recall. The action involved a software anomaly that could trigger Error Code 45 (EC45), causing an unexpected pump shutdown. Affected units were those manufactured or updated between May 2007 and February 2011. A pump shutdown during active therapy could result in a delay or interruption of infusion, which the FDA identified as carrying the potential for serious injury or death. Moog notified affected customers and provided corrective software updates. At the time of the announcement, no adverse patient events had been reported to the company.
Facilities operating Curlin 6000 CMS pumps should confirm the software version and service history of any unit in their fleet to verify that the EC45 correction has been applied. IPR Medical can inspect and service Curlin 6000 CMS units and advise on software status during repair engagements.
Action recommended: Facilities using Curlin 6000 CMS pumps manufactured or updated between May 2007 and February 2011 should verify that the EC45 software correction has been applied before continued use. Contact IPR Medical at 855-477-8866 or cs@iprmed.com for assistance with device inspection and service.
Because FDA device corrections and safety communications may be updated or expanded after the date of this listing, IPR Medical recommends that facilities also check the FDA’s official resources directly to ensure they have the most current information:
IPR Medical — Your Curlin 6000 CMS Service Partner
IPR Medical buys, sells, repairs, and stocks parts and administration sets for the Moog Curlin 6000 CMS. Whether you’re looking to add verified, service-ready units to your fleet, send in a pump for preventive maintenance or corrective repair, or simply keep your consumable supply stocked, our team has the expertise and inventory to support you. We work with home infusion providers, hospitals, long-term care facilities, and independent biomedical departments across the country.
IPR Medical buys, sells, repairs, and stocks parts and IV sets for the Moog Curlin 6000 CMS. Contact us at 855-477-8866 or cs@iprmed.com.
| Weight | 1 lbs |
|---|---|
| Dimensions | 2.5 × 4 × 5.1 in |
| Condition | Refurbished |
- AC Power Adapter
- Bolus/Patient Cable
- Lock Box (500 mL)
- Pole Clamp
- Holster