The McGaw BD 360 syringe pump offers continuous intravenous infusion rates up to 360 mL/hr, depending on syringe size. It functions as both a primary infusor and can administer downstream secondary piggyback infusions. The pump’s digital controls allow for precise variable time settings, ensuring accurate delivery of medication. An LCD display provides clear visibility of the remaining infusion time, enhancing monitoring and management. Its versatility makes it suitable for various medical scenarios. The BD 360’s advanced features support efficient and reliable infusion therapy. It is an essential tool for healthcare providers seeking precision and ease of use. Overall, the McGaw BD 360 syringe pump is a robust solution for intravenous infusion needs.

Send In BD McGaw 360 Infuser for Repair
  • Dimensions: 9.75” H x 2.75” W x 5” D
  • Weight: 2 lbs
  • Syringes: BD Plastipak™ 3cc to 60cc
  • Accuracy: ± 3%
  • Battery: Requires 4 C-size alkaline batteries
  • Battery Life: 400 hours typical, 10 hours on Low Battery
  • Displays:
      • LED: Infuse, Attention, Low Battery, Infusion Complete.
      • LCD: Time Remaining, Original Set Time; some revisions have Infusion complete indicator.

    Occlusion Limits (using BD Plastipak™ syringes):

    Syringe

    Size

    Maximum Occlusion

    Pressure (psi)

    Typical Occlusion

    Pressure (psi)

    3 cc 46.8 36.4
    5 cc 30.0 17.9
    10 cc 22.8 17.0
    20 cc 21.6 13.9
    30 cc 16.8 9.4
    60 cc 13.2 8.4
  • Rate Parameters: 1.5 mL/hr (3cc syringe) to 360 mL/hr (60cc syringe)
  • Delivery Time: 10 – 60 min., in 2.5 min. increments

BD McGaw 360 Infuser

So here you will be able to find the dimensions, weight, syringes and rate parameters of BD McGaw 360 Infuser Infusion Pump Repair .

 

 

  • Pole Clamp