Philips M2734A / M2734B Smart Toco MP Transducer | Avalon
The Philips M2734A and M2734B are wired Smart Toco MP transducers used with the Philips Avalon family of fetal/maternal monitors (FM20, FM30, FM40, FM50). They combine two parameters in a single sensor — external uterine activity (Toco) and maternal pulse (MP) — and enable the Avalon platform’s Smart Pulse automated coincidence detection, which cross-checks the maternal pulse against the fetal heart rate to confirm the host monitor is not inadvertently tracking maternal heart rate as fetal. The M2734B is the successor revision to the M2734A and is functionally interchangeable on the Avalon Toco socket; both are auto-identifying via the Smart-interface (pink) connector and carry internal calibration data storage.
| Device Type | Wired dual-parameter Smart Toco MP transducer (uterine activity + maternal pulse) |
| Models | M2734A (original Smart Toco MP) and M2734B (successor revision); both share the same role and Avalon socket |
| Compatible Monitors | Philips Avalon FM20 (M2702A), FM30 (M2703A), FM40 (M2704A), FM50 (M2705A) at supported software revisions |
| Connector | Wired Smart-interface (pink) connector to the dedicated Toco socket on the Avalon host monitor |
| Key Features | Dual-parameter Toco + maternal pulse, Smart Pulse coincidence detection, waterproof housing for shower/bath use, latex-free, auto-identification, internal calibration data storage |
| Patient Population | Antepartum testing and labor and delivery |
IPR Medical buys, sells, repairs, and stocks parts for the Philips Avalon fetal monitor family, including the M2734A and M2734B Smart Toco MP transducers. Contact us at 855-477-8866 or cs@iprmed.com.
Philips M2734A / M2734B Smart Toco MP Transducer — Specifications
| Manufacturer | Philips Healthcare |
| Model Numbers | M2734A (original Smart Toco MP) and M2734B (successor revision); also referenced as Philips part 453564833171 |
| Device Type | Wired dual-parameter Smart Toco MP transducer for non-invasive monitoring of uterine activity and maternal pulse |
| Intended Use | External, non-invasive monitoring of uterine activity (duration, frequency, and relative intensity of contractions) and simultaneous maternal pulse acquisition; supports Avalon Smart Pulse cross-channel verification of FHR vs. maternal pulse |
| Toco Sensor | Pressure-sensitive resistor bridge (DMS / strain-gauge element) with AC excitation; output amplified and synchronously rectified to a pressure-proportional DC signal proportional to abdominal-wall hardness |
| Maternal Pulse Sensor | Integrated maternal pulse acquisition within the same housing; signal forwarded to the Avalon host monitor for Smart Pulse processing alongside the Toco channel |
| Compatible Host Monitors | Philips Avalon FM20 (M2702A), Avalon FM30 (M2703A), Avalon FM40 (M2704A), and Avalon FM50 (M2705A) fetal/maternal monitors |
| Connector | Philips Smart-interface (pink) connector that mates with the dedicated Toco socket on the front face of the Avalon host monitor |
| Auto-Identification | Transducer self-identifies to the Avalon monitor on connection; the host monitor automatically configures the Toco channel and Smart Pulse function without operator action |
| Internal Calibration | On-board memory stores calibration data so the transducer travels with its calibration profile across host monitors |
| Smart Pulse / Coincidence Detection | Automated cross-channel verification on supported Avalon configurations: the host monitor compares the maternal pulse from this transducer to the FHR from the paired ultrasound transducer and alerts the clinician when the two coincide, helping rule out misidentification of maternal heart rate as fetal |
| Baseline / Zeroing | Toco baseline key on the host monitor zeroes the Toco display and trace to 20 units (relative units) when monitoring externally; coupling gel is not required for the Toco channel |
| Belt Attachment | Snap-clip belt fixing onto compatible reusable washable fetal belts (e.g., M1562A) or waterproof washable fetal belts (e.g., M1562B); also compatible with disposable abdominal belts (e.g., M2208A) |
| Housing | Waterproof housing intended to support patient mobility during labor, including shower/bath use; latex-free patient contact materials |
| Coupling Medium | No coupling gel required (Toco/MP transducers operate by mechanical pressure and integrated MP sensing, not ultrasound) |
| Cleaning & Disinfection | Surface clean and disinfect per the Avalon FM20/30 or FM40/50 Instructions for Use; do not steam-sterilize |
| Compatible Accessories | Reusable washable fetal belts M1562A; waterproof washable fetal belts M1562B; disposable abdominal belts M2208A |
| Reference Documentation | Philips Avalon Fetal Monitor FM20/FM30 and FM40/FM50 Instructions for Use and corresponding Service Guides |
| Regulatory Classification | Accessory to a Class II fetal cardiotocograph under 21 CFR 884.2660 (FDA Obstetrics/Gynecology panel); host monitor cleared via 510(k) by the original manufacturer |
| FDA Device Corrections | FDA Class 2 device recall records exist for both the M2734A (Recall Event ID 175943) and the M2734B (Recall Event ID 176765) in the FDA Medical Device Recalls Database. See the Description Tab for further detail and links to the FDA’s official records. |
IPR Medical buys, sells, repairs, and stocks parts for the Philips Avalon fetal monitor family, including the M2734A and M2734B Smart Toco MP transducers. Contact IPR Medical at 855-477-8866 or cs@iprmed.com with service or software questions.
Trademark Notice: Philips, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Philips.
Overview
The Philips M2734A and M2734B are the OEM wired Smart Toco MP transducers designed for use with the Philips Avalon family of fetal/maternal monitors — the FM20 (M2702A), FM30 (M2703A), FM40 (M2704A), and FM50 (M2705A). Unlike a conventional external Toco transducer, which measures only uterine activity, the Smart Toco MP combines two parameters in a single sensor: external uterine activity from a strain-gauge pressure element and maternal pulse from an integrated MP sensor. That dual-parameter design is what enables the Avalon platform’s Smart Pulse automated coincidence detection — a feature unique to the Avalon FM series at its introduction — which compares the maternal pulse from this transducer against the fetal heart rate from the paired ultrasound transducer and alerts the clinician when the two coincide. This helps rule out one of the most consequential errors in external CTG monitoring: a host monitor that is silently tracking maternal heart rate while clinicians believe they are watching the fetus. The M2734B is the successor revision to the M2734A and is functionally interchangeable on the Avalon Toco socket.
Connector, Auto-Identification, and Calibration Memory
The M2734A and M2734B terminate in a Philips Smart-interface (pink) connector that mates with the dedicated Toco socket on the front face of the Avalon host monitor. On connection, the transducer self-identifies to the host monitor over the Smart-interface, and the monitor automatically configures the Toco channel, enables maternal pulse acquisition, and activates Smart Pulse coincidence detection if the host configuration supports it. The transducer carries on-board memory that stores its calibration data, so the calibration profile travels with the transducer rather than being held by the host monitor. This means a transducer can be moved between Avalon monitors in a Labor and Delivery unit without recalibration at each bedside, and a host monitor can accept any compatible Smart-interface transducer without local pairing steps. Auto-identification also reduces operator setup error: the host monitor cannot be tricked into displaying Toco data on the wrong channel because it reads the transducer’s identity electronically rather than relying on socket assignment alone.
Smart Pulse and Cross-Channel Verification
External fetal monitoring carries a known clinical risk: the ultrasound Doppler beam can lock onto maternal vessels (typically the maternal aorta or major pelvic arteries) and produce a heart-rate trace that looks plausible but is actually maternal. If the maternal pulse and fetal pulse coincide for an extended period, clinicians can be falsely reassured that the fetus is well when in reality the FHR signal is being lost or has degenerated. The Smart Pulse feature on the Avalon FM series uses the maternal pulse channel of the M2734A/M2734B to perform automated coincidence detection: the host monitor continuously compares the maternal pulse measured from the Smart Toco MP transducer to the FHR measured from the paired ultrasound transducer, and when the two readings overlap closely enough to suggest the FHR may actually be maternal, the host monitor alerts the clinician on screen. Importantly, this cross-channel verification works without requiring the patient to wear a maternal SpO₂ sensor or maternal ECG electrodes, simplifying setup and reducing the cabling at the bedside while preserving the clinical safety benefit.
Strain-Gauge Toco Operation and Baseline Behavior
The Toco channel of the M2734A/M2734B uses the same fundamental sensing approach as a conventional external Toco transducer. As the uterus tightens during a contraction, the abdominal wall stiffens and pushes outward against the Smart Toco MP; inside the housing, that mechanical force deflects a movable base plate against a pressure-sensitive resistor bridge — a strain-gauge (DMS) element. AC excitation is applied to the bridge, and the resulting differential output is amplified and synchronously rectified to produce a pressure-proportional DC signal that is forwarded to the host monitor for display on the Toco channel and printing on the trace. Because the signal is relative rather than calibrated absolute pressure, the Toco baseline key on the Avalon host monitor is used to zero the trace to 20 units between contractions. Belt tightness, transducer placement over the fundus, and patient body habitus all affect signal amplitude, so brief repositioning of the transducer or re-tensioning of the belt is normal during a study. The MP channel runs in parallel with the Toco channel inside the same housing and does not require coupling gel.
Compatibility with the Philips Avalon Family
The M2734A and M2734B are designed for use with the Avalon FM20 (M2702A), Avalon FM30 (M2703A), Avalon FM40 (M2704A), and Avalon FM50 (M2705A) fetal/maternal monitors. All four monitors share the same wired Smart-interface ecosystem, the same Toco socket, and the same Smart Pulse software architecture, so the Smart Toco MP transducer can be moved between any of the four host monitor types within a Labor and Delivery unit. The host monitor must be at a software revision that supports Smart Pulse and the Smart-interface transducer family; older Avalon software builds may identify the transducer and operate the Toco channel correctly but may not present the full coincidence-detection feature set. Facilities running mixed software revisions across an Avalon fleet should verify with their biomed team or with IPR Medical that the host monitor’s software level supports all the Smart Toco MP features they expect to use. The M2734A and M2734B are functionally interchangeable; the M2734B is the successor revision and may be supplied in place of the M2734A for service replacement.
Patient Mobility, Belts, and Workflow
The Smart Toco MP attaches to a fetal monitoring belt with a snap-clip belt fixing identical in interface to other wired Avalon transducers, so existing belts in a Labor and Delivery unit’s circulating stock continue to work. Compatible belts include reusable washable fetal belts (M1562A), waterproof washable fetal belts (M1562B) for showering and tub labor, and disposable abdominal belts (M2208A) for single-patient use. The transducer housing is waterproof, so it can remain in place during showering and bathing without removing the patient from continuous CTG and maternal pulse monitoring, provided the patient remains within reach of the host monitor’s cable. After the belt is in place, the operator palpates for the fundus, positions the Smart Toco MP over that area, snaps it onto the belt, tensions the belt to a comfortable but firm fit, and presses the Toco baseline key on the monitor to zero the trace between contractions. The host monitor automatically begins displaying maternal pulse on the configured channel without additional operator action. After each patient, the transducer is wiped clean of skin oils and lotions, surface-disinfected per the host system’s care and cleaning instructions, and never steam-sterilized.
Service, Testing, and Preventive Maintenance
The Avalon FM20/30 and FM40/50 service guides define structured parameter tests that biomedical staff use to validate Smart Toco MP function in the field. The host monitor includes built-in self-tests, and the dedicated Toco parameter test exercises the front-end Toco circuitry together with an attached transducer to confirm correct strain-gauge response and baseline behavior; the maternal pulse channel is verified separately. Common bench failures on aged Smart Toco MP transducers include drifting Toco baseline that cannot be zeroed (typically pointing to strain-gauge degradation or moisture ingress past a compromised housing seal), intermittent maternal pulse acquisition (often related to the MP sensor or its internal cabling), cable flex damage near the strain relief and at the Smart-interface connector backshell, and connector pin damage from repeated insertion. The Smart-interface connector and its on-board calibration memory are integral to the assembly; component-level repair of either is generally not recommended, and the standard remedy for a failed transducer is replacement of the assembly. Performance verification after replacement should follow the same parameter test sequence the host system’s service guide defines for the Toco and MP channels.
Service and Maintenance Considerations
The M2734A and M2734B are wired transducers with on-board calibration memory rather than networked devices, and they have no software, cybersecurity, or EMR integration of their own; data transmission, alerts, trend storage, and integration with obstetrical surveillance systems such as Philips OB TraceVue and IntelliSpace Perinatal are handled by the host Avalon monitor. From a clinical engineering perspective the transducer should be inspected at each preventive maintenance interval for housing cracks, wear of the patient-contact base plate, cable jacket integrity, strain-relief integrity at both the transducer body and the Smart-interface connector, and connector pin condition. Because the Smart Toco MP has on-board calibration memory, a transducer that is functioning correctly should display consistent Toco baseline behavior across any Avalon host monitor it is plugged into; significant drift on one host monitor but not another typically points to a host-monitor issue rather than the transducer. IPR Medical’s biomed team can support facilities that need bench testing, evaluation, or replacement of M2734A and M2734B transducers as part of fleet maintenance on Avalon fetal monitors.
FDA Device Corrections and Safety Bulletins
FDA Class 2 medical device recall records exist in the FDA Medical Device Recalls Database for both revisions of this transducer: the M2734A under Recall Event ID 175943 and the M2734B under Recall Event ID 176765. Class 2 indicates a situation in which use of, or exposure to, the device may cause temporary or medically reversible adverse health consequences, and the probability of serious adverse health consequences is remote. IPR Medical recommends that clinical engineering teams review the FDA’s official records for both recalls directly to confirm the affected lot or date ranges, the recalling firm’s reference and reason, the specific corrective action requested, and the current status of each recall, because these details may be updated by FDA after the date of this listing. Facilities operating Philips Avalon fetal monitors should also review the broader Philips fetal-monitoring transducer safety communications referenced on the Philips customer support portal, including any field correction or service notices that affect transducers refurbished beyond their useful life.
Because FDA device corrections and safety communications may be updated or expanded after the date of this listing, IPR Medical recommends that facilities also check the FDA’s official resources directly to ensure they have the most current information.
FDA resources for further review:
FDA Medical Device Recalls Database |
FDA Medical Device Safety
IPR Medical — Buy, Sell, Repair, and Parts
IPR Medical buys, sells, repairs, and stocks parts and accessories for the Philips Avalon fetal monitor family, including OEM M2734A and M2734B Smart Toco MP transducers. Whether your facility needs a single replacement Smart Toco MP to bring an Avalon FM20, FM30, FM40, or FM50 back into service, additional units to expand monitoring across more Labor and Delivery beds, or evaluation and repair of an existing transducer that is producing baseline drift, intermittent maternal pulse acquisition, or Smart-interface auto-identification faults, IPR Medical’s biomed team can support the request. We also buy back surplus Avalon monitors and Smart-interface transducers from facilities upgrading their fetal monitoring fleet, and can supply matching ultrasound transducers, cordless transducers, patient modules, fetal belts, and recorder paper to keep existing units in service. Reach out for revision verification (M2734A vs. M2734B), lot/serial verification against current FDA recall scope, fleet sourcing for multi-bed deployments, and service or software questions on the host monitor.
IPR Medical buys, sells, repairs, and stocks parts for the Philips Avalon fetal monitor family, including the M2734A and M2734B Smart Toco MP transducers. Contact IPR Medical at 855-477-8866 or cs@iprmed.com with service or software questions.
Trademark Notice: Philips, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Philips.
| Product Condition |
|---|
Compatible Parts and Accessories
IPR Medical stocks a range of compatible parts and accessories for the Philips Avalon fetal monitor family on which the M2734A and M2734B Smart Toco MP transducers are used. Contact us at 855-477-8866 or cs@iprmed.com to confirm availability, pricing, or compatibility for your specific configuration.
Wired Smart-Interface Transducers (Avalon)
| Smart Toco MP Transducer (M2734A) | Original revision of the wired Smart Toco MP for Avalon FM20/FM30/FM40/FM50; dual-parameter Toco + maternal pulse with Smart Pulse coincidence detection. |
| Smart Toco MP Transducer (M2734B) | Successor revision to the M2734A, functionally interchangeable; supplied as the standard service replacement (Philips part 453564833171). |
| Smart Ultrasound Transducer (M2735A / M2735B) | Companion wired Smart-interface ultrasound transducer for fetal heart rate; used together with the Smart Toco MP to enable Smart Pulse coincidence detection. |
| Smart Ultrasound Transducer (M2736A / M2736B) | Additional Smart-interface ultrasound transducer revision for FHR monitoring on Avalon hosts. |
Cordless (Wireless) Transducers — Avalon CTS
| Avalon CTS Cordless Fetal Transducer System (M2720A) | Cordless transducer base station that pairs with the M2725A/M2726A/M2727A wireless transducers and connects to the Avalon FM20/FM30/FM40/FM50 host monitor (alternative to the wired Smart-interface chain). |
| Cordless Toco Transducer (M2725A / 862488) | Wireless Toco transducer for the Avalon CTS chain; cordless alternative to the wired Smart Toco MP for ambulatory monitoring. |
| Cordless Ultrasound Transducer (M2726A / 862489) | Wireless ultrasound transducer for fetal heart rate via the Avalon CTS base station. |
| Cordless ECG Transducer (M2727A / 862490) | Wireless ECG transducer for direct fetal ECG and maternal ECG (depending on host monitor configuration) via the Avalon CTS base station. |
Compatible Avalon Host Monitors
| Avalon FM20 (M2702A) | Antepartum fetal/maternal monitor; supports Smart Toco MP via the M2734A and M2734B on the Toco socket. |
| Avalon FM30 (M2703A) | Antepartum and intrapartum fetal/maternal monitor with internal monitoring options; full Smart Pulse support at supported software revisions. |
| Avalon FM40 (M2704A) | Antepartum fetal/maternal monitor in alternate form factor; full Smart Pulse support at supported software revisions. |
| Avalon FM50 (M2705A) | Antepartum and intrapartum fetal/maternal monitor with internal monitoring options; full Smart Pulse support at supported software revisions. |
Belts and Belt Accessories
| Reusable Washable Fetal Belts (M1562A / 989803104111) | Standard reusable abdominal belts for Avalon transducers; supplied in packs of 3. |
| Waterproof Washable Fetal Belts (M1562B / 989803129891) | Reusable waterproof belts intended for use during shower/bath labor with the waterproof Smart Toco MP; supplied in packs of 3. |
| Disposable Abdominal Belts (M2208A / 989803106021) | Single-patient-use disposable abdominal belts for Avalon transducers; supplied in cases of 100. |
DECG / MECG Cables, Patient Modules, and Electrodes
| DECG Adapter Cable (M1362B / 989803103551) | Adapter cable for direct fetal ECG monitoring on FM30 / FM50 monitors when paired with a compatible patient module. |
| MECG Adapter Cable (M1363A / 989803103561) | Adapter cable for maternal ECG monitoring on Avalon monitors; for use with M1364A or M1365A patient modules. |
| DECG/MECG Patient Module (M1364A / M1365A) | Patient-worn module that accepts DECG legplate adapter cable M1362B and MECG adapter cable M1363A on Avalon FM30 / FM50. |
| Disposable Spiral Scalp Electrodes | Single-spiral and double-spiral fetal scalp electrodes for use with the M1357A DECG legplate on FM30 / FM50. Contact IPR Medical for the specific OEM part number matching your region. |
Recorder Paper and Consumables
| Avalon Trace Paper | OEM thermal Z-fold trace paper for the Avalon FM20/FM30 and FM40/FM50 integrated recorders; scaling and form factor vary by monitor model and region. Contact IPR Medical for the specific OEM part number matching your Avalon model and region. |
IPR Medical buys, sells, repairs, and stocks parts for the Philips Avalon fetal monitor family, including the M2734A and M2734B Smart Toco MP transducers. Contact IPR Medical at 855-477-8866 or cs@iprmed.com with service or software questions.