Baxter Mini-Infuser 300XL Syringe Pump | Multi-Speed
The Baxter Mini-Infuser 300XL is a battery-powered, multi-speed syringe infusion pump originally developed by Bard Medsystems and later distributed by Baxter Healthcare for controlled, intermittent intravenous medication administration. The 300XL adds a three-position rate selector switch to the same chassis used by the single-rate 150XL, allowing the clinician to choose among three plunger drive speeds and infuse standard 5–60 mL disposable syringes over a wider range of durations without changing syringe sizes. It is well suited to short-duration secondary infusions, IV piggyback delivery, and small-volume drug administration in inpatient, outpatient, and ambulatory care settings.
| Pump Type | Multi-speed (multispeed) syringe infusion pump |
| Compatible Syringes | 5 mL to 60 mL disposable syringes (B-D, Monoject) |
| Plunger Drive Rates | XLow 1.8 in/hr, Low 2.7 in/hr, Medium 5.5 in/hr (rate selector switch) |
| Power | Four C-size alkaline cells; ~400 hours typical service life |
| Portability | 1.75 lbs with batteries; pole loop and optional mounting bracket |
| Notable Features | Active time scale, anti-syphon latch, occlusion detection, audible/visual alarms |
IPR Medical buys, sells, repairs, and stocks parts for the Baxter Mini-Infuser 300XL. Contact us at 855-477-8866 or cs@iprmed.com.
Also available from IPR Medical:
Baxter Mini-Infuser 150XL — single-rate sister pump on the same chassis |
Baxter Auto Syringe AS40A — programmable syringe pump for continuous and intermittent infusion
Baxter Mini-Infuser 300XL — Specifications
| Manufacturer | Baxter Healthcare Corporation (originally Bard Medsystems Division, C.R. Bard, Inc.) |
| Model | Mini-Infuser 300XL |
| Product Codes | 2M8171 |
| Device Classification | Multi-speed (multispeed) syringe infusion pump |
| Intended Use | Controlled intermittent IV medication administration via heparin lock, primary line piggyback, or volume-control chamber piggyback |
| Pump Type | Linear-drive syringe pump with three selectable plunger speeds |
| Plunger Drive Rates | XLow 1.8 in/hr, Low 2.7 in/hr, Medium 5.5 in/hr; selected via rate selector switch on the back of the case; delivered fluid flow rate varies with syringe diameter |
| Compatible Syringes | 5 mL to 60 mL disposable syringes; B-D and Monoject sizes 5, 10, 20, 30, 35, and 60 mL |
| Accuracy | ±3% (not including syringe tolerance) |
| Flow Profile | Continuous, flat (constant rate at the selected speed) |
| Active Time Scale | Active time scale on the case face indicates infusion duration based on selected rate and syringe size |
| Occlusion Detection Time | 70 seconds |
| Occlusion Force | 8.0 ± 1.0 lbs |
| Volume Stored on Occlusion | Approximately 9 mL |
| Back Pressure Effect on Accuracy | None to occlusion pressure |
| Anti-Syphon Protection | Mechanical anti-syphon latch captures syringe plunger flange |
| Alarms and Indicators | Flashing green (normal operation), flashing red (end of syringe or occlusion), flashing yellow (low battery), audible alarm in “On with Alarm” mode |
| Power | Four “C” size alkaline cells |
| Battery Life | Approximately 400 hours typical service; approximately 4 hours after low-battery indication |
| Low Battery Threshold | 4.0 ± 0.2 volts |
| Dimensions | 5 in (H) × 4.5 in (W) × 1.6 in (D) |
| Weight | 1.75 lbs (1 lb 12 oz) with batteries installed |
| Mounting | Integrated IV pole loop; optional mounting bracket for bedrails, headboards, and other surfaces |
| Case Material | Impact-resistant ABS molded case |
| Connectivity | None — standalone mechanical device, no network or data interface |
| Drug Library | Not applicable — multi-speed mechanical pump without programmable drug library |
| FDA 510(k) Clearance | K930459, cleared August 10, 1993, applicant Bard Medsystems Division (cleared as “Bard Mini-Infuser 1500 & 3000,” marketed as 150XL/300XL); product code FRN; 21 CFR 880.5725; Device Class II |
| Manufacturer Service Status | Baxter has ended factory service on this product; third-party service and parts support remain available through IPR Medical |
| FDA Device Corrections | Class 3 recall Z-2269-2012 (initiated March 4, 2011; terminated August 27, 2012) addresses C-cell battery contact compatibility. See the Description Tab for full detail. |
IPR Medical buys, sells, repairs, and stocks parts for the Baxter Mini-Infuser 300XL. Contact IPR Medical at 855-477-8866 or cs@iprmed.com with service or software questions.
Trademark Notice: Baxter, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Baxter.
Overview
The Baxter Mini-Infuser 300XL is a compact, lightweight, battery-powered linear syringe pump designed for the controlled intermittent administration of intravenous medications. The pump was originally developed by the Bard Medsystems Division of C.R. Bard, Inc. and first cleared by the FDA in 1993; it was later marketed by Baxter Healthcare Corporation as the Mini-Infuser 150XL (single-rate) and Mini-Infuser 300XL (multi-speed). The 300XL holds a standard 5 mL, 10 mL, 20 mL, 30 mL, 35 mL, or 60 mL disposable syringe in a barrel clamp and advances the plunger at one of three selectable linear speeds — XLow 1.8 inches per hour, Low 2.7 inches per hour, or Medium 5.5 inches per hour — chosen via a rate selector switch on the back of the case. The selected plunger drive rate combined with the syringe diameter determines the fluid flow rate, allowing the clinician to achieve a specific infusion duration without changing the syringe size. The 300XL is well suited to short-duration medication infusions in critical care, oncology, infectious disease, neonatal, pediatric, and ambulatory care settings, particularly where the broader rate range expands the practical infusion durations achievable from a single syringe.
Display, Indicators, and User Interface
The Mini-Infuser 300XL uses a simplified mechanical interface rather than a programmable display, which keeps the operating workflow short and reduces training overhead for facilities that need a reliable single-purpose device. Operation centers on a three-position switch (Off, On, and On with Alarm), a release lever that disengages the pusher and anti-syphon latch for syringe loading and unloading, and a rate selector switch on the back of the case that chooses among the XLow, Low, and Medium plunger drive speeds. Three colored LED indicators communicate pump status: a flashing green light confirms normal operation, a flashing red light signals end-of-syringe or an occlusion, and a flashing yellow light indicates a low-battery condition. In the On with Alarm position, an audible beeping alarm mimics the flashing red light to provide both visual and acoustic notification. The 300XL also features an active time scale and indicating mark on the case face that shifts with the rate selection, so the clinician can read remaining infusion time directly without consulting a separate nomogram every time the rate is changed.
Drug Delivery and Infusion Configurations
The Mini-Infuser System is designed around three primary administration configurations. The first configuration uses the pump as an intermittent infusion system in conjunction with a heparin lock, with a Microbore extension set connecting the syringe to the patient access. The second configuration is a piggyback delivery in combination with a gravity primary IV line, where the Microbore extension set is attached to the lower injection port or flashball site of the primary set so that the primary flow continues uninterrupted while the syringe contents are delivered. The third configuration is a piggyback in combination with a volume-control chamber such as a Buretrol set, which allows further dilution of the syringe contents into the burette before delivery. On the 300XL specifically, the rate selector lets the clinician extend the same syringe over a longer infusion period at XLow or Low, or run a faster delivery at Medium when the protocol calls for shorter administration. Drugs are typically diluted to a final concentration in the syringe such that one syringe delivery equals one prescribed dose at the desired duration; the Baxter operator’s manual provides nomograms for each compatible B-D and Monoject syringe so clinicians can quickly determine fill volume and total minutes-to-complete for any rate-and-syringe combination.
Multi-Speed Mechanical Design and 300XL Versus 150XL
The Mini-Infuser product family includes two pumps that share the same chassis, syringe holder, and electrical platform but differ in their delivery flexibility. The 300XL is a multi-speed pump with a rate selector switch on the back of the case that selects among three plunger drive speeds — XLow at 1.8 inches per hour, Low at 2.7 inches per hour, and Medium at 5.5 inches per hour — expanding the practical infusion duration range without changing syringe sizes. Because the ratio between XLow and Medium is approximately three to one, the operator’s manual notes that minutes-to-complete at XLow is roughly three times the corresponding value at Medium for the same syringe. The 150XL is the single-rate sister product, intended for facilities that want the simplest possible workflow for a single fixed plunger drive rate. Because the 300XL is a fixed-mechanical-rate pump, it does not maintain a programmable drug library, dose error reduction software, or any electronic settings beyond the on/off/alarm switch and rate selector. The same battery, syringe holder, anti-syphon latch, and alarm logic are shared between the two models, which simplifies parts inventory and biomed training when a facility carries both. IPR Medical stocks both the 150XL and the 300XL and can advise on which model best matches a specific clinical workflow when replacements are being considered.
Battery Operation, Mounting, and Portability
The Mini-Infuser 300XL is exclusively battery operated and does not include an AC power option. Four standard “C” size alkaline cells provide approximately 400 hours of typical service life, and the pump alarms with a flashing yellow light at a battery threshold of 4.0 ± 0.2 volts; once the low-battery indication appears, approximately 4 hours of remaining infusion time is typical before the pump can no longer drive the plunger. Battery selection is operationally significant because the FDA-classified Class 3 correction in 2011 (described below) traced inoperability to a contact-diameter incompatibility between the pump’s battery spring and certain over-the-counter C cells, so consistent sourcing of compatible alkaline cells matters for reliability. The pump weighs 1.75 lbs with batteries installed and measures 5 inches tall by 4.5 inches wide by 1.6 inches deep, which makes it convenient for ambulatory and bedside use. Mounting options include an integrated IV pole loop molded into the case and an optional mounting bracket that secures the pump to bedrails, headboards, transport carts, and other surfaces. The impact-resistant ABS molded enclosure is appropriate for routine clinical handling, though the manual notes the pump should not be immersed and is not suitable for use in the presence of flammable anesthetics.
Service, Maintenance, and Calibration Considerations
Baxter has ended factory service on the Mini-Infuser 300XL, which means hospitals that continue to operate these pumps must rely on third-party service organizations or in-house biomedical engineering teams for repair, calibration, and parts. The Baxter operator’s manual specifies a regular calibration check that includes verifying free movement of the syringe holder mechanism, confirming proper retention by the anti-syphon latch, validating the audible alarm, and confirming that downward force at the syringe holder of approximately 8.0 ± 1.0 lbs triggers the occlusion alarm within the specified 70-second detection window. Service technicians should also verify correct rate selection at each of the three plunger drive speeds. The manual recommends lubricating the lead screw at least every six months using the manufacturer-supplied grease and applicator. Because the 300XL has no electronic display, no event log, and no networked diagnostic interface, biomed inspections rely on direct mechanical and electrical checks rather than on software-based service screens. IPR Medical performs service, calibration, and repair on the Mini-Infuser 300XL and can supply replacement components such as the pole clamp, mounting bracket, and battery cover when these are confirmed available from OEM or compatible-parts sources.
FDA Device Corrections and Safety Bulletins
The FDA Medical Device Recalls Database lists one device correction for the Mini-Infuser 300XL: a Class 3 recall, FDA recall number Z-2269-2012, initiated by Baxter Healthcare Corporation on March 4, 2011 and terminated on August 27, 2012. The correction is paired with the parallel 150XL recall (Z-2268-2012) and addresses a battery contact compatibility issue. The Mini-Infuser battery spring has a flat contact measuring approximately 0.53 inches on the inner diameter of the spring coil; the IEC 60086-2 and ANSI 08 standards specify a minimum contact diameter for C-cell batteries of 0.51 inches. If a C cell with the minimum allowable contact diameter is installed, the contact circuit may not complete and the Mini-Infuser will not power on. The same battery interface design is shared with the Baxter InfusO.R. pump, and Baxter expanded the original InfusO.R. safety alert (issued January 11, 2011) to include the Mini-Infuser 150XL and 300XL with the March 2011 correction. The recall covers all serial numbers of product codes 2M8171 (new) and 2M8171R (refurbished). Approximately 19,380 units of the 300XL were affected, with a distribution pattern covering roughly 45,865 units across the combined Mini-Infuser System. The FDA classified the action as Class 3, indicating that use of the affected product is not likely to cause adverse health consequences, because the failure mode is a clear and immediate failure to power on rather than a delivery accuracy or alarm failure during active infusion.
Because FDA device corrections and safety communications may be updated or expanded after the date of this listing, IPR Medical recommends that facilities also check the FDA’s official resources directly to ensure they have the most current information.
FDA resources for further review:
FDA Medical Device Recalls Database |
FDA Medical Device Safety
IPR Medical — Buy, Sell, Repair, and Parts
IPR Medical is an independent secondary-market specialist in the Baxter Mini-Infuser 300XL. Because Baxter has ended factory service on this product, hospitals, ambulatory clinics, veterinary practices, and infusion service organizations that continue to use the 300XL rely on third-party support for replacement units, repair, calibration, and parts. IPR Medical purchases retired Mini-Infuser 300XL pumps for fleet refresh, sells recertified 300XL units that have been tested against OEM-defined functional and accuracy criteria across all three plunger drive speeds, repairs and recalibrates customer-owned pumps, and stocks the most commonly requested wear and accessory parts including pole clamps and mounting hardware. Recertified pumps include the standard rubber boot, pole clamp, and operator’s guide. Whether a facility needs to consolidate its multi-speed Mini-Infuser fleet, recover capital by selling pumps no longer in clinical rotation, or extend the service life of existing units through repair, IPR Medical provides a single point of contact for the full product life cycle.
IPR Medical buys, sells, repairs, and stocks parts for the Baxter Mini-Infuser 300XL. Contact IPR Medical at 855-477-8866 or cs@iprmed.com with service or software questions.
Trademark Notice: Baxter, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Baxter.
| Weight | 5 lbs |
|---|---|
| Dimensions | 14 × 10 × 10 in |
Compatible Parts and Accessories
IPR Medical stocks a range of compatible parts and accessories for the Baxter Mini-Infuser 300XL. Contact us at 855-477-8866 or cs@iprmed.com to confirm availability, pricing, or compatibility for your specific configuration.
Administration and Extension Sets
| Microbore IV Extension Set | Small-bore extension tubing used between the syringe and the patient access (heparin lock, primary line injection port, or volume-control chamber); recommended in the Baxter operator’s manual to minimize bolus volume in the event of an occlusion. Contact IPR Medical for the specific OEM part number matching your length and connector configuration. |
Compatible Disposable Syringes
| B-D Plastic Syringes, 5–60 mL | Becton Dickinson disposable plastic syringes in 5, 10, 20, 30, and 60 mL sizes are listed by Baxter as compatible with the Mini-Infuser 300XL barrel clamp and flange notch. |
| Monoject Disposable Syringes, 5–60 mL | Sherwood/Covidien Monoject syringes in 5, 6, 12, 20, 35, and 60 mL sizes are listed in the Baxter operator’s manual nomograms and accepted by the syringe holder. |
Power and Battery
| C-Size Alkaline Cells (set of 4) | Four “C” size alkaline batteries are required to power the Mini-Infuser 300XL. Per FDA recall Z-2269-2012, facilities should standardize on a brand verified to power on reliably with the Mini-Infuser battery spring; contact IPR Medical for guidance on confirmed-compatible C-cell sourcing. |
| Battery Cover | Replacement battery compartment cover for the Mini-Infuser 300XL case. Contact IPR Medical for the specific OEM part number matching your case revision. |
Mounting and Transport Accessories
| Pole Clamp Assembly | Pole clamp accessory for IV pole mounting; included as standard with IPR Medical recertified Mini-Infuser 300XL pumps. Contact IPR Medical for the specific OEM part number for replacement units. |
| Optional Mounting Bracket | Mounting bracket for securing the pump to bedrails, headboards, transport carts, and other surfaces, as described in the Baxter operator’s manual. Contact IPR Medical for the specific OEM part number matching your mounting configuration. |
| Protective Rubber Boot | Protective rubber case boot for impact resistance during clinical handling and transport; included as standard with IPR Medical recertified Mini-Infuser 300XL pumps. |
Service and Documentation
| Lead Screw Lubrication Grease and Applicator | Manufacturer-supplied grease and applicator used for the Baxter-recommended six-month lead screw lubrication cycle; required to maintain accurate plunger drive over the service life of the pump. |
| Operator’s Manual, Mini-Infuser 150XL/300XL | Baxter Mini-Infuser 150XL and 300XL operator’s manual, including syringe nomograms for B-D and Monoject syringes and the FDA correction reference letter. |
IPR Medical buys, sells, repairs, and stocks parts for the Baxter Mini-Infuser 300XL. Contact IPR Medical at 855-477-8866 or cs@iprmed.com with service or software questions.
Reviews
There are no reviews yet.