Stryker/Gaymar T/Pump TP700
The Stryker/Gaymar T/Pump® TP700 is a professional-grade localized temperature therapy system that circulates temperature-controlled water through a Stryker Mul•T•Pad® to deliver precise warming or cooling therapy to targeted body areas. Four selectable setpoints — 50°F (cooling), 95°F, 100°F, and 107°F — cover everything from post-operative hypothermia prevention and edema management to chronic joint pain and muscle strain relief. The TP700 upgrades over the classic TP600 with a fourth warming setpoint (95°F), programmable therapy cycles (20-minute, 30-minute, or continuous), and a setpoint lockout feature that prevents accidental temperature changes mid-treatment. A three-layer safety architecture — dual temperature sensors, dual microprocessors, and a mechanical over-temperature thermostat — ensures the pad never exceeds the 107°F maximum contact surface temperature. At 6.5 lbs empty and with built-in hose and cord management, the TP700 is easy to position at bedside, in recovery, or in the OR.
IPR Medical buys, sells, and repairs the Stryker/Gaymar T/Pump TP700 and stocks compatible Mul•T•Pad® accessories and replacement parts. Contact us at 855-477-8866 or cs@iprmed.com.
Only 1 left in stock
Send In Stryker/Gaymar T/Pump TP700 for RepairStryker/Gaymar T/Pump® TP700 — Specifications
| Manufacturer | Stryker Medical (formerly Gaymar Industries) |
| Model | T/Pump® TP700 (also TP700C with quick-disconnect connectors) |
| Device Type | Localized temperature therapy system — warming and cooling |
| Temperature Setpoints | 107°F (42°C) | 100°F (38°C) | 95°F (35°C) | 50°F (10°C — cooling with ice) |
| Temperature Accuracy | ±2°F at 107°F (±1°C at 42°C) |
| Maximum Contact Surface Temperature | 107°F (42°C) |
| High Limit Safety Temperature | 110°F–120°F (43.3°C–49°C) — mechanical thermostat shuts off heater |
| Therapy Cycle Options | Continuous, 30-minute, or 20-minute (TP700 series only) |
| Setpoint Lockout | Yes — press and hold 2 seconds to lock or unlock (TP700 series only) |
| Safety System | Three-layer: dual thermistor sensors, dual microprocessors, mechanical over-temperature thermostat |
| Flow Rate | 7 gph (26.5 lph) minimum with pad attached |
| Reservoir Capacity | 93 oz (2.75 l) maximum |
| Hose Length | 10 ft (305 cm) dual hose, attached |
| Connector Type | Clik-Tite® (TP700) or quick-disconnect (TP700C) |
| Compatible Pads | Stryker Mul•T•Pad® (multiple sizes: 3″×23″, 13″×18″, 15″×22″, 18″×26″); single-use and reusable options |
| Electrical Requirements (120V) | 120 VAC ±10%, 60 Hz, 3.1 A |
| Electrical Requirements (230V) | 230 VAC +10%, 50 Hz, 1.6 A |
| Electrical Classification | Class I equipment, Type BF applied part, continuous operation |
| Certifications | UL 60601-1, CAN/CSA C22.2 No. 601.1, ASTM F 2196-2002, EN 60601-1, EN 60601-1-2, IEC 60601-1 |
| Dimensions (approx.) | 11.5″ × 8″ × 8″ (29.2 cm × 20.3 cm × 20.3 cm) |
| Weight (empty) | 6.5 lbs (2.9 kg) | 9 lbs (4.0 kg) filled to heating level |
| Ambient Operating Temperature | 60°F–90°F (15.6°C–32.2°C) |
| Transport / Storage Temperature | -20°F to 120°F (-28°C to 48°C), uncontrolled RH |
| Current Leakage | 300 microamperes maximum (120V model) |
| Ground Resistance | 0.5 ohm maximum |
| FDA Device Corrections | Class 3 recall Z-1013-2018 (Stryker Medical, posted March 2018) covering TP700 and TP700C models. See the Descriptions for details and FDA resources. |
IPR Medical buys, sells, and repairs the Stryker/Gaymar T/Pump TP700 and stocks compatible Mul•T•Pad® accessories and replacement parts. Contact us at 855-477-8866 or cs@iprmed.com.
Trademark Notice: Stryker, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Stryker.
Overview
The Stryker/Gaymar T/Pump® TP700 is a professional localized temperature therapy system designed for clinical use in hospitals, surgical suites, recovery rooms, orthopedic clinics, and long-term care settings. The system circulates temperature-regulated water from an internal reservoir through a Stryker Mul•T•Pad® applied directly to the patient’s skin, delivering consistent conductive thermal therapy without the variability of ice packs or electric heating pads. The TP700 is the professional-tier model in the T/Pump line, distinguished from the classic TP600 by its four temperature setpoints, programmable therapy cycle timers (20-minute, 30-minute, or continuous), and an operator-controllable setpoint lock that prevents inadvertent changes during active therapy.
Warming and Cooling Therapy
Heat therapy promotes vasodilation and increased blood flow to the treated area, making the TP700 effective for chronic joint pain, arthritis, tendonitis, muscle strain, post-operative recovery, and prevention of hypothermia during and after surgery. The three warming setpoints — 95°F, 100°F, and 107°F — provide clinical flexibility: the 95°F setting supports gentle warming for sensitive patients, while the 107°F setting is appropriate for robust heat therapy where tolerability has been established. Cooling therapy at 50°F assists in vasoconstriction, reduces inflammation, and provides pain relief during the acute phase of injury. The cooling mode uses ice added to the reservoir; as ice melts and the setpoint begins flashing, the operator refills and restarts the cycle. Multiple Mul•T•Pads can be interconnected via Clik-Tite® connectors for simultaneous therapy at more than one body site from a single pump unit.
Three-Layer Safety System
Patient safety is the engineering core of the TP700. The system uses two independent thermistor sensors feeding two separate microprocessors — one for temperature control and one dedicated to over-temperature detection. If the control processor fails to shut off the heater before the water temperature reaches the high-limit threshold (110°F–120°F), a third independent layer — a mechanical thermostat — interrupts the heater circuit and stops heat delivery. This redundant architecture prevents a single-point electronics failure from causing a dangerous over-temperature event. Maximum contact surface temperature is capped at 107°F regardless of setpoint selection, and the flow indicator cuts heater power if the pump is tipped and water flow stops. Stryker’s guidance calls for checking the patient’s skin every 30 minutes during active therapy and applying additional vigilance for patients who are unconscious, diabetic, or have impaired circulation.
Therapy Cycles and Setpoint Lock
The TP700’s programmable therapy cycle feature allows caregivers to select a 20-minute, 30-minute, or continuous therapy duration directly from the keypad. Timed cycles allow the system to run unattended for a defined treatment period without requiring staff to return and manually shut off the unit. The setpoint lock — unique to the TP700 series — prevents the patient or an unauthorized user from changing the prescribed temperature. Pressing and holding the lock button for two seconds engages or disengages the lock; the selected temperature remains active even if the keypad is accidentally bumped. This feature is particularly valuable in unsupervised recovery and pain management settings where consistent temperature delivery is critical.
Pads and Accessories
The TP700 is designed for exclusive use with Stryker Mul•T•Pad® products. The Mul•T•Pad’s random-flow internal button design allows water to circulate evenly and permits the pad to be folded into a custom fit around joints, limbs, or the torso without kinking the flow path — reducing the need for multiple specialty pad sizes in a facility’s inventory. Pads are available in four sizes (3″×23″, 13″×18″, 15″×22″, and 18″×26″) in both single-patient-use and reusable polymer variants. The TP700 ships with a 10-foot dual hose that connects to the pad via Clik-Tite® connectors (or quick-disconnect on the TP700C model), with built-in hose and cord storage on the pump body for organized transport. IPR Medical stocks Mul•T•Pad accessories and replacement parts for the TP700 platform.
FDA Device Corrections and Safety Bulletins
The FDA database records one device recall for the Stryker T/Pump TP700 and TP700C: recall number Z-1013-2018, a Class 3 action posted by the FDA on March 14, 2018 (Stryker Medical Division of Stryker Corporation, FEI #1831750). A Class 3 recall is the FDA’s least severe recall classification, defined as a situation in which use of the product is unlikely to cause adverse health consequences. Facilities with TP700 or TP700C units in service should verify their specific unit serial numbers against the FDA recall database to confirm applicability. IPR Medical recommends contacting Stryker Medical customer service at 1-800-327-0770 for unit-specific guidance.
Recommended actions for facilities with TP700 or TP700C units: Check your unit’s serial number against the FDA Medical Device Recalls Database using the links below. Because FDA device corrections and safety communications may be updated or expanded after the date of this listing, IPR Medical recommends that facilities also check the FDA’s official resources directly to ensure they have the most current information.
Official FDA resources for the most current information:
• FDA Medical Device Recalls Database
IPR Medical — Sales, Repair, and Parts
IPR Medical buys, sells, and repairs the Stryker/Gaymar T/Pump TP700 and stocks compatible Mul•T•Pad® accessories and replacement parts. Whether you need a refurbished unit, a bench repair on an existing pump, or replacement pads and hose assemblies, IPR Medical can support your facility from a single source. Contact us for multi-unit pricing or fleet service inquiries.
IPR Medical buys, sells, and repairs the Stryker/Gaymar T/Pump TP700 and stocks compatible accessories and replacement parts. Contact us at 855-477-8866 or cs@iprmed.com.
Trademark Notice: Stryker, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Stryker.
| Weight | 11 lbs |
|---|---|
| Dimensions | 11.5 × 10.2 × 12.5 in |
Compatible Parts and Accessories
IPR Medical stocks a range of compatible parts and accessories for the Stryker/Gaymar T/Pump® TP700. The sections below cover the most commonly requested components, sourced from the Stryker T/Pump TP600 & TP700 Series Service Manual (Document 101176 REV E) and the TP700/TP700C Operations Manual (Document 101515 REV E). Contact us at 855-477-8866 or cs@iprmed.com to confirm availability, pricing, or compatibility for your specific configuration.
Control Units
| TP700 | T/Pump® Professional Control Unit with Clik-Tite® connectors — four temperature setpoints, programmable therapy cycles, setpoint lock |
| TP700C | T/Pump® Professional Control Unit with quick-disconnect connectors — same features as TP700; specified for use with TP6xxE quick-disconnect pad series |
Mul•T•Pad® — Clik-Tite® Connector Pads (for use with TP700)
The following single-patient-use and reusable pads connect to the TP700 via Clik-Tite® connectors. All are compatible with the TP700’s 10-foot dual hose. Multiple pads may be interconnected for simultaneous multi-site therapy.
| TP22B | Mul•T•Pad® 15″w × 22″l (38 cm × 56 cm) — All polymer construction; 10 per carton |
| TP22C | Mul•T•Pad® 15″w × 22″l (38 cm × 56 cm) — Nonwoven fabric one side, pliable polymer other side; single unit per carton |
| TP3E | Mul•T•Pad® 3″w × 23″l (8 cm × 58 cm) — Nonwoven fabric one side, pliable polymer other side; narrow wrap/limb pad; 10 per carton |
| TP12E | Mul•T•Pad® 13″w × 18″l (33 cm × 46 cm) — Nonwoven fabric one side, pliable polymer other side; 20 per carton |
| TP22E | Mul•T•Pad® 15″w × 22″l (38 cm × 56 cm) — Nonwoven fabric one side, pliable polymer other side; 20 per carton |
| TP26E | Mul•T•Pad® 18″w × 26″l (46 cm × 66 cm) — Nonwoven fabric one side, pliable polymer other side; largest standard pad; 10 per carton |
| TP22G | Mul•T•Pad® 15″w × 22″l (38 cm × 56 cm) — Heavy polymer, reusable construction; 10 per carton |
Mul•T•Pad® — Quick-Disconnect Pads (for use with TP700C)
The following pads use quick-disconnect fittings and are specified for use with the TP700C control unit. They are not compatible with standard Clik-Tite® connectors.
| TP612E | Mul•T•Pad® 13″w × 18″l (33 cm × 46 cm) — Nonwoven fabric on each side with quick-disconnect fittings; 10 per carton |
| TP622E | Mul•T•Pad® 15″w × 22″l (38 cm × 56 cm) — Nonwoven fabric on each side with quick-disconnect fittings; 20 per carton |
| TP626E | Mul•T•Pad® 18″w × 26″l (46 cm × 66 cm) — Nonwoven fabric on each side with quick-disconnect fittings; 10 per carton |
Power and Electrical Components
| Hospital Grade Power Cord — 120V | 3-wire hospital-grade power cord for 120 VAC / 60 Hz operation; must connect to a properly grounded hospital-grade receptacle |
| Hospital Grade Power Cord — 230V | Modular 3-wire cordset for 230 VAC / 50 Hz international operation; plug configuration may require local conversion adapter |
Hose and Connection Hardware
| Attached Dual Hose Assembly (Clik-Tite®) | 10 ft (305 cm) dual-channel hose with Clik-Tite® connectors; attaches pad to pump; compatible with TP700 and all Clik-Tite® Mul•T•Pad models |
| Attached Dual Hose Assembly (Quick-Disconnect) | 10 ft (305 cm) dual-channel hose with quick-disconnect fittings; compatible with TP700C and TP6xxE quick-disconnect Mul•T•Pad series |
| Hose Pinch Clamps | Inline hose clamps used to open and close water flow to pad; serrated-end push-to-open design; used during pad connection, disconnection, and storage procedures |
| Tube Set Strap | Retention strap for securing coiled hose to pump body during storage and transport |
Reservoir and Fill Components
| Tethered Fill Cap | Quarter-turn tethered reservoir cap; prevents cap loss; seals the 93 oz (2.75 l) water reservoir; standard on all TP700 series units |
IPR Medical buys, sells, and repairs the Stryker/Gaymar T/Pump TP700 and stocks compatible Mul•T•Pad accessories and replacement parts. Contact IPR Medical at 855-477-8866 or cs@iprmed.com to confirm availability, pricing, or compatibility for your specific unit configuration.
Reviews
There are no reviews yet.