Quick answer

Moog Medical publishes the current Infinity Enteral Feeding Pump Operator’s Manual (LS-25356-001 Rev. T, 2025), and the copy we host below is the same edition. The Infinity Orange, the small-volume model, has its own manual (LS-28945-001 Rev. J, 2025) on Moog’s website. The standard (Blue) Infinity delivers 0.1 to 600 ml/hr and the Infinity Orange 0.1 to 300 ml/hr, both with ±5% accuracy, a lithium-ion battery rated for 24 hours at 125 ml/hr, and no routine calibration requirement. Use the manual that matches your model, and check the software revision the pump shows at start-up.

Whether you are training caregivers, checking an alarm at the bedside or setting up an equipment program for a home care fleet, the right Infinity document saves time. This guide collects the documents we host, points to the current manufacturer editions, explains which one you need, and pulls out the specifications people ask us about most.

Download the Enteralite Infinity Manuals

Infinity Operator’s Manual, LS-25356-001 Rev. T (current)

The full English and French operator’s manual for the standard Infinity Enteral Feeding Pump (the Blue model), copyright 2025 Moog Inc.: warnings, components, priming and loading, single and interval feeding, battery, user preference settings, alarms, troubleshooting, cleaning, accessories and specifications.

Download PDF

Infinity Orange Operator’s Manual, LS-28945-001 Rev. J (current)

The English and French operator’s manual for the Infinity Orange small-volume enteral feeding pump, copyright 2025 Moog Inc., hosted by Moog Medical. Use this edition for Orange pumps; its rate range, sets and keypad differ from the standard Infinity.

View on moogmedical.com

Infinity Product Brochure, MD-41258-001 Rev. E

Moog’s 2023 eight-page brochure for the standard Infinity: headline features, specifications, ordering numbers for the pump kit, delivery sets, charger, pole clamp and carry packs.

Download PDF

Moog Medical Download Library

Moog also publishes quick reference instructions, troubleshooting guides, patient and family guides and delivery set instructions for both the Infinity and the Infinity Orange, in several languages.

View on moogmedical.com

These documents are provided for reference. They are the original manufacturer’s documents and are not maintained or updated by IPR Medical. For more manuals, browse our Manuals & Documentation Library.

Which Enteralite Infinity Manual Do You Need?

The pump started life as the ZEVEX EnteraLite Infinity. Moog Inc. acquired ZEVEX in 2007, and Moog Medical now publishes the manuals under the Infinity name. There are two pump models, and each has its own manual:

DocumentWho uses itNotes
Infinity Operator’s Manual (LS-25356-001 Rev. T)Clinicians, caregivers, educators, biomedFor the standard Infinity (Blue). The keypad has a FEED INT key for interval feeding. The copy we host matches Moog’s current download.
Infinity Orange Operator’s Manual (LS-28945-001 Rev. J)NICU and pediatric staff, caregivers, biomedFor the small-volume Infinity Orange. The keypad has a FOOD TYPE key (formula or human milk) instead of FEED INT.
Quick reference instructions and troubleshooting guidesPatients, families and cliniciansSummaries only, available from Moog for each model; not a substitute for the full manual.
Product brochure (MD-41258-001 Rev. E)Purchasing and clinical evaluationSpecifications and order numbers for the standard Infinity.

Moog does not list a service manual in its public download library, and the operator’s manuals state that the pump has no user-serviceable parts. At start-up the pump shows its nine-digit serial number, then the software revision: the letter R followed by a number on the standard Infinity, and the letter N followed by a number on the Infinity Orange. Not sure which model or revision you have? Contact us. For recall and correction history by serial number, see our Enteralite Infinity recalls and safety notices guide.

What’s Inside the Operator’s Manual

  • Warnings and cautions: approved sets only, set change intervals, battery capacity, electrical safety during cleaning, homemade blenderized foods, and line entanglement.
  • Pump components: door, pump wheel, upstream and downstream pressure sensors, air sensor, cassette receptacle, keypad, display and symbols.
  • Directions for use: priming and loading the set, a single feeding and an interval feeding example, infinite dose, and the battery.
  • User preference settings: alarm volume, settings lock, dose done alarm, backlight and auto prime (the Orange adds maximum rate and maximum dose limits).
  • Alarms, messages and indications: every display message, with why it happens and how to correct it, plus a troubleshooting flowchart.
  • Cleaning, accessories and specifications: acceptable and unacceptable cleaners, pole clamps and carry packs, and the full specification table.
  • Pediatric considerations and technical information: accuracy conditions, head height effect, set residual volume and electromagnetic compatibility.

Key Specifications: Infinity and Infinity Orange

From Moog’s current Operator’s Manuals (Infinity LS-25356-001 Rev. T and Infinity Orange LS-28945-001 Rev. J) and the Infinity brochure (MD-41258-001 Rev. E). Where the documents differ, both values are shown.

SpecificationInfinity (Blue)Infinity Orange
Flow rate range0.1 to 600 ml/hr0.1 to 300 ml/hr
Rate increments0.1 ml/hr from 0.1 to 10 ml/hr; 1 ml/hr from 10 to 600 ml/hr0.1 ml/hr from 0.1 to 10 ml/hr; 1 ml/hr from 10 to 300 ml/hr
Accuracy±5% (or 0.050 ml/hr, whichever is greater, per the pediatric section) with Infinity sets±5% or 0.050 ml/hr, whichever is greater
Dose range0.1 to 3000 ml, or infinite dose0.1 to 3000 ml
Special featuresInterval feeding (FEED INT)Food type setting (formula or human milk) and optional maximum rate and dose limits
PrimingHold to prime or auto prime, about 700 ml/hrHold to prime or auto prime, about 360 ml/hr
Set volume when primed / residualAbout 15 ml / about 11.5 mlAbout 5 ml / about 1.5 ml
Occlusion alarmsNO FLOW IN −5 psi ±3; NO FLOW OUT 12 psi ±3NO FLOW IN −5 psi ±3; NO FLOW OUT 12 psi ±3 (30 s delay), 15 psi ±2 (5 s delay) or 18 psi ±2 (1 s delay)
Downstream occlusion time to alarm5 minutes at 125 ml/hr; 2 hours at 1 ml/hr; 12 hours at 0.1 ml/hr. Up to 1 ml is released after the occlusion clears.
Maximum infusion pressure40 psi dead head
BatteryLithium-ion; 24 hours at 125 ml/hr, 9 hours at 600 ml/hrLithium-ion; 24 hours at 125 ml/hr, 9 hours at 300 ml/hr
Charge time and battery service lifeAbout 6 hours from fully depleted; charge at 10 to 40 °C; expected battery service life 2 to 5 years depending on use
ChargerInput 100–240 VAC, 50–60 Hz, 400 mA; output 5 VDC, 3.0 A; charger rated IP42, electrical protection Class II
Size1.95 in D × 5.65 in W × 4.05 in H (4.9 × 14.4 × 10.3 cm)
Weight14.4 oz; 411.0 g in the manual, 408 g in the brochure14.4 oz (411.0 g)
Water protection and dropIP25 (water jets from any direction); designed so a drop from 3 feet does not damage pump operation; operates in any orientation
Operating conditions5 to 40 °C (41 to 104 °F); 10 to 95% humidity, non-condensing; 700 to 1,060 hPa
Storage conditions−20 to 50 °C (−4 to 122 °F); 10 to 95% humidity, non-condensing
MaintenanceNo routine calibration or adjustment required; no user-serviceable parts
Service lifeDesigned for a minimum of 5 years of service life

Delivery Sets

The manuals warn to use only Infinity brand disposable sets; others will not deliver the correct dose, may allow dangerous free-flow conditions and may generate hazardous fluid pressures that set off occlusion alarms at unpredictable pressures. The current sets use ENFit connectors.

  • Infinity (Blue): Infinity 500 ml Bag Set (INF0500-E), Infinity 1200 ml Bag Set (INF1200-E) and Infinity Safety Screw Spike Set (INF0020-E).
  • Infinity Orange: Infinity Orange Screw Cap Set (INFOCAP-E) and Infinity Orange 100 ml Bag Set (INF0100-E), plus the INF0500-E, INF1200-E and INF0020-E sets.
  • Important: Moog’s brochure notes that Infinity Orange delivery sets cannot be used with the standard Infinity pump.
  • Change intervals: replace bag sets every 24 hours and spike sets every 48 hours (maximum 1.5 L), and do not program rate and dose combinations that run past those limits.

Quick Reference from the Manual

TopicWhat the manual says
Start-up self-testThe pump shows its serial number three digits at a time, then the software revision, then lights every display segment for 2 seconds; verify all segments and symbols are active.
First use and chargingCharge for 6 hours before running on battery. If the pump shuts down on low battery, charging less than 10 minutes may cause an ER99 alarm.
Low batteryLOW BATT flashes every 3 seconds with a beep every 2 seconds when 1 hour or less remains.
Self-test errorsER01 – ER99: turn the pump off and back on, then check all settings and volume counters before feeding.
Air detectionNO FOOD sounds when about 1 ml of air (about 5 inches of tubing) enters the teal tubing (Infinity).
Food type (Orange)Turning the pump off and on resets food type to formula; if dose and food type are not viewed before RUN/PAUSE, the pump displays CON FIRM DOSE AND FOOD TYPE.
CleaningWarm soapy water, a 5% bleach and water solution or Metrex CaviCide are acceptable; alcohol, Ball Ruthless, Wex-Cide 128 and Vesphene IIse are not. Rinse under warm water and dry. Never clean with the pump on or the charger plugged in.
Hanging the bagHang the bag so its bottom is at or above the level of the pump door; head height of ±12 inches changes accuracy by about ±2.5%.

For what these messages usually mean in practice, see our Enteralite Infinity common problems and repair guide.

Service and Repair

IPR Medical evaluates, repairs and tests Enteralite Infinity and Infinity Orange pumps in-house at our ISO 9001:2015-certified facility in Tustin, California. U.S. customers get a free inbound shipping label and a free evaluation, and nothing is charged until you approve the estimate. Learn more on our Moog pump repair service page, or shop new and recertified (refurbished) Enteralite Infinity Blue and recertified Infinity Orange feeding pumps.

Enteralite Infinity Manual FAQ

Where can I download the Enteralite Infinity operator’s manual?

You can download the current Infinity Operator’s Manual (LS-25356-001 Rev. T) from this page. Moog Medical publishes the same edition, and the Infinity Orange manual (LS-28945-001 Rev. J), in its online download library.

What is the difference between the Infinity and the Infinity Orange?

The standard (Blue) Infinity delivers 0.1 to 600 ml/hr and supports interval feeding. The Infinity Orange is a small-volume pump for 0.1 to 300 ml/hr with a food type setting for formula or human milk, smaller sets with about 1.5 ml of residual volume, and optional maximum rate and dose limits. Infinity Orange sets cannot be used on the standard Infinity.

What is the flow rate range of the Enteralite Infinity?

The standard Infinity runs from 0.1 to 600 ml/hr, in 0.1 ml/hr steps up to 10 ml/hr and 1 ml/hr steps above that, with ±5% accuracy using Infinity sets. The Infinity Orange runs from 0.1 to 300 ml/hr.

How long does the Infinity battery last?

The manual rates a full charge for 24 hours at 125 ml/hr (9 hours at 600 ml/hr on the standard Infinity, 9 hours at 300 ml/hr on the Orange), with about 6 hours to recharge. The battery’s expected service life is 2 to 5 years depending on use.

Does the Enteralite Infinity need calibration?

Moog’s manuals state that no routine calibration or adjustment is required and that there are no user-serviceable parts. Many facilities still include the pump in a yearly performance check under their own equipment management program.

How do I find the software revision on an Infinity pump?

Turn the pump on and watch the self-test. After the nine-digit serial number, the display shows the software revision: the letter R and a number on the standard Infinity, or the letter N and a number on the Infinity Orange.

Is the Enteralite Infinity waterproof?

It is water resistant, not waterproof. The pump is rated IP25, and the manual allows rinsing it under a stream of warm water after cleaning, with the pump off and the charger unplugged. The AC adapter/charger is rated only IP42 and must be completely dry before it is plugged in.

Need Help With an Enteralite Infinity?

Questions about a manual, an alarm or a repair? Our team can help, Monday through Friday, 7:00 AM to 4:30 PM Pacific.