Skip to content
IPR Medical
Infusomat Space Version G
Infusion PumpsManuals & DocumentationB. Braun Infusomat Space
Quick answer

B. Braun hosts the current U.S. Infusomat Space Instructions for Use for Software U (valid for software 586U) on bbraunusa.com, and you can download an earlier 586U edition, the Infusomat Space pump set catalog and the Space system brochure below. Always match the manual to the software version shown on your pump’s start-up screen. The Infusomat Space delivers 0.1 to 1200 mL/h with ±5% rate accuracy, offers nine downstream occlusion levels from 225 to 900 mmHg, runs on a NiMH or Li-Ion battery pack, and B. Braun recommends a documented Technical Safety Check every two years.

Whether you are training new staff, checking an alarm at the bedside or setting up a biomed PM program, the right Infusomat Space document saves time. This guide collects the documents we host, explains which one you need, and pulls out the specifications and reference details people ask us about most.

Download the Infusomat Space Manuals

Instructions for Use, Software U (current)

The U.S. Instructions for Use that B. Braun hosts on its Infusomat Space Large Volume Pump, Wireless product page, valid for software 586U. Back-page code 38911910, revision 1025 (88 pages). Use this edition for pumps on U software.

View on bbraunusa.com

Instructions for Use, Software 586U (earlier edition)

A 90-page U.S. edition for software 586U (back-page code 38911910, revision 0321): line loading, drug library programming, primary and secondary infusions, alarms, wireless drug library upload, AutoProgramming, battery care, technical data, cleaning and accessories.

Download PDF

Infusomat Space Pump Sets Catalog

B. Braun’s four-page 2010 listing of Infusomat Space pump sets with ULTRASITE and SAFELINE needle-free systems, with catalog numbers, drop factors, priming volumes and lengths.

Download PDF

Space Infusion System Brochure

B. Braun’s 2016 brochure covering the Infusomat Space (2nd generation) and Perfusor Space pumps, the SpaceStation, KeyGuard keypad and wireless interoperability.

Download PDF

These documents are provided for reference. They are the original manufacturer’s documents and are not maintained or updated by IPR Medical. For more manuals, browse our Manuals & Documentation Library.

Which Infusomat Space Manual Do You Need?

The Instructions for Use state on the cover which software they are valid for, and recommend that all pumps at your institution run the same software version. The pump shows its software version during the start-up self-test, and you can also find it in the Status menu. Here is how the documents fit together:

DocumentWho uses itNotes
Instructions for Use (by software version)Clinicians, educators, biomedBoth editions linked above are valid for software 586U (Version U). Pumps on earlier 686-series software (Versions E, F and G) need the edition written for that software; request it from B. Braun.
Pump sets catalogClinicians, materials managementThe pump is intended for use only with dedicated Infusomat Space disposables. The 2010 catalog is a starting point; confirm current catalog numbers with B. Braun.
Service manual and HiBaSeD service program instructionsBiomedical techniciansThe Instructions for Use refer to the HiBaSeD service program for pump logs, configuration and calibration, and say service manuals and service training are available from B. Braun. Not hosted here.
SpaceStation, SpaceStation MRI and Battery Pack SP with WiFi instructionsBiomed, ITSeparate documents for the docking station, MRI use and wireless battery module.

Not sure which software your pump runs, or whether a recall applies? Our Infusomat Space recall and software versions guide explains how to check and what the FDA-tracked notices mean.

What’s Inside the Instructions for Use

  • Overview and patient safety: controls and LEDs, pole clamp mounting and stacking, indications, warnings, and the MRI and fluid-ingress cautions.
  • Operation: loading and priming the line, drug library programming, secondary (piggyback) infusions, intermittent doses, loading doses, boluses, care unit changes, line changes and standby.
  • Pump menus: downstream and upstream pressure, alarm volume, dose rate calculator, data lock, KVO, display settings, bolus rate and wireless activation, plus the Status menu.
  • Alarms: device alarms, pre-alarms, operating alarms, reminder alarms and prompts.
  • Wireless drug library upload and AutoProgramming: receiving drug libraries and EHR orders over the network.
  • Battery operation and maintenance: charging, storage, the WiFi battery, wireless status codes and the battery maintenance cycle.
  • Technical data, service, cleaning and accessories: specifications, trumpet curves, the Technical Safety Check, cleaning and disinfecting, and ordering numbers.

Key Infusomat Space Specifications

From the Infusomat Space Instructions for Use for software 586U and the Space system brochure. Confirm against the manual for your pump’s software version.

SpecificationInfusomat Space
TypeVolumetric infusion pump for adults, pediatrics and neonates
Rate range0.1 to 1200 mL/h; increments of 0.01 mL/h below 100 mL/h, 0.1 mL/h from 100 to 999.9 mL/h, 1 mL/h above 1000 mL/h
Volume to be infused0.1 to 9999 mL; time 00:01 to 99:59 h
Rate accuracy±5% per IEC/EN 60601-2-24; bolus typically ±5% for volumes over 1 mL
Set change intervalPumping accuracy maintained for at least 96 hours
Downstream occlusion9 levels, 225 to 900 mmHg; maximum post-occlusion bolus 2 mL under fault conditions
Upstream sensor9 levels, -90 to -160 mmHg
Air detectionUltrasonic sensor; single bubble alarm 0.02 to 0.3 mL, accumulated air 0.5 to 3.8 mL/h (both configurable)
Drug libraryUp to 1200 drug names, 10 concentrations per drug, 30 categories, 50 care units and 16 patient profiles
Battery run timeLi-Ion: 4.0 h at 100 mL/h and 2.5 h at 1200 mL/h with wireless active; 12.0 h and 5.0 h with wireless inactive. NiMH: 13 h at 100 mL/h, 5 h at 1200 mL/h
Recharging timeApproximately 6 hours
Power100–240 V AC, 50/60 Hz via the SpaceStation or Power Supply SP; 11–16 V DC external low voltage
Wireless802.11 a/b/g/n with the Battery Pack SP with WiFi (8713182U)
Size and weight214 × 68 × 124 mm (8.4 × 2.6 × 4.8 in); approximately 1.4 kg
Fluid protectionIP22 (drip protected for horizontal use)
ClassificationFDA Class II (21 CFR 880.5725, product codes FRN and FPA); defibrillator-proof, Type CF; Protective Class II (Class I in combination with the SpaceStation)
Operating conditions15 to 40 °C; 30 to 90% relative humidity, non-condensing
MRIUnsafe as a standalone device near MRI equipment; MR Conditional only within the SpaceStation MRI
Technical Safety CheckRecommended every 2 years

Ordering numbers in the Instructions for Use: 8713051U (pump with Battery Pack SP with WiFi, Li-Ion) and 8713052U (pump with Battery Pack SP without WiFi, NiMH).

Infusomat Space Pump Sets

The Instructions for Use say the Infusomat Space is intended for use only with dedicated disposables labeled for each route of administration, and that the technical data were determined with the Infusomat Space line 870 0036 SP. The manual lists these set families:

  • 15 drop sets
  • 60 drop sets
  • Blood sets
  • Epidural sets
  • Filtered sets

The 2010 pump set catalog adds burette, add-on, low adsorption and UV-resistant sets with ULTRASITE or SAFELINE injection sites. Several Infusomat Space administration set lots and catalog numbers have been the subject of FDA recalls, including corrections for false air-in-line alarms and backcheck valve backflow; see our recall and software versions guide before you rely on older set stock.

Quick Reference from the Manual

TopicWhat the manual says
Software versionDisplayed with Self-test active at power-up, and in the Status menu.
Device alarmSwitch the pump off and on; if the device alarm repeats, the pump must be sent for service.
Battery storageBefore storage of more than 5 months, fully charge the battery pack and remove it from the pump.
Battery maintenanceThe pump asks for a maintenance cycle every 30 days when enabled; the full cycle takes about 12 hours.
Battery lifeApproximately 500 charge cycles before lifetime decreases; a symbol blinking for more than 1 hour on AC means the battery should be checked.
Safety checkTechnical Safety Check recommended every 2 years and documented; a dropped pump must be checked by service.
CleaningEarlier 586U edition we host: mild soapy water, then disinfectants containing 1-propanol, isopropyl alcohol, ethanol or didecyl dimethyl ammonium chloride; no chloride (bleach) products. Current B. Braun-hosted edition: adds diisobutylphenoxyethyl dimethyl benzyl ammonium chloride and sodium hypochlorite, but still no bleach on the peristaltic fingers and membrane. Never spray disinfectant directly on the pump; the earlier edition adds to let it dry at least 20 minutes.
ConnectorsKeep fluid out of the P2 and P3 connectors; cover P2 with an unplugged power supply or combi lead plug while cleaning, then inspect for moisture and plating damage.
MountingUp to three Space pumps per pole clamp; do not position the pump over the patient or lean on it; do not use a damaged pole clamp.
Stuck doorUse the crank key stored in the battery compartment cover to remove the emergency aperture cover and turn the door open.

For what these messages usually mean in practice, see our Infusomat Space common problems and repair guide.

Service and Repair

IPR Medical evaluates, repairs and tests Infusomat Space pumps on Versions F, G and U in-house at our ISO 9001:2015-certified facility in Tustin, California. U.S. customers get a free inbound shipping label and a free evaluation, and nothing is charged until you approve the estimate. Every repair carries a 90-day warranty. We update pump software as part of service; ask us about version upgrades. Learn more on our B. Braun infusion pump repair page, or shop recertified (refurbished) Infusomat Space Version U and Version G pumps.

Infusomat Space Manual FAQ

Where can I download the Infusomat Space Instructions for Use?

B. Braun publishes the current U.S. Instructions for Use for Software U (586U) on its website, linked above. You can also download an earlier 586U edition, the pump set catalog and the Space system brochure from this page.

How do I find the software version on an Infusomat Space?

The software version is displayed with the Self-test active message when you turn the pump on. You can also open the Status menu from the Home screen and scroll to the version.

What is the flow rate range of the Infusomat Space?

0.1 to 1200 mL/h, with ±5% rate accuracy according to IEC/EN 60601-2-24. Facilities can also configure lower maximum limits for continuous and bolus rates in the pump configuration file.

How long does the Infusomat Space battery last?

The 586U Instructions for Use list 4.0 hours at 100 mL/h for the Li-Ion pack with wireless active, 12.0 hours with wireless inactive, and 13 hours for the NiMH pack. Recharging takes approximately 6 hours.

Can the Infusomat Space be used in MRI?

Not on its own. The manual says the pump is unsafe as a standalone device near MRI equipment and can be used conditionally in the MR environment only within the SpaceStation MRI, following its instructions.

Can I use bleach to clean an Infusomat Space?

It depends on the edition. The earlier 586U edition we host says not to use chloride (bleach) products. The current B. Braun-hosted edition lists sodium hypochlorite as an approved disinfectant for the housing but still prohibits bleach on the peristaltic fingers and membrane. Follow the edition that matches your pumps and B. Braun’s current guidance.

How often does an Infusomat Space need a safety check?

B. Braun recommends a documented Technical Safety Check every two years, and the manual requires a pump that has been dropped or exposed to force to be checked by service. If you need a pump serviced, request a free shipping label and we will evaluate it.

Need Help With an Infusomat Space?

Questions about a manual, an alarm or a repair? Our team can help, Monday through Friday, 7:00 AM to 4:30 PM Pacific, or contact us online.