Version 8.2 Sigma Spectrum
Infusion PumpsFDA Recalls & SoftwareSigma Spectrum V6 and V8
Quick answer

We found 25 FDA-tracked recall and correction events involving Sigma Spectrum V6 or V8 pumps, their drug library software or their battery modules since 2010: eight Class I, sixteen Class II and one Class III. The most recent Class I recall, initiated in June 2025 (Z-2095-2025 for V6 and Z-2096-2025 for V8), covers specific pumps that may have the wrong platform’s software installed, so check that V6 software begins with 6 and V8 software begins with 8. The most recent event overall, in November 2025 (Z-1121-2026), covers certain V8 pumps released from service with defective gear grease. Two older Class I corrections still shape daily practice: inspect rear case pins and battery contacts for residue and corrosion (2020), and on V8 pumps, never press RUN/STOP before resolving an upstream occlusion (2021).

The Sigma Spectrum has a long FDA record. Some notices involve software and are fixed by an update or reversion; others involve hardware, labeling or pumps released from service without full testing. This guide puts the V6 and V8 notices in one place, in plain English, so your team can tell which pumps are affected and what to do next.

Who Makes the Sigma Spectrum

The Spectrum was developed by SIGMA International General Medical Apparatus, LLC of Medina, New York, which is why older pumps, manuals and recall notices carry the SIGMA name. In 2009, Baxter signed an exclusive global distribution agreement with SIGMA and took a 40 percent equity stake with an option to buy the rest. Baxter exercised that option and completed the purchase in April 2012. Baxter Healthcare Corporation is the current manufacturer and the recalling firm on recent notices. Spectrum IQ, with Dose IQ Safety Software, is a separate Baxter pump; several notices below also covered it, but this guide covers only the V6 and V8 parts of them.

Sigma Spectrum Software Versions

V6 and V8 pumps look alike, and the software version is what tells you which notices apply. These are the versions named in the manual we host and in FDA records:

Platform and versionStatusNotes
4.02.06OlderThe v6.05 manual says yellow keyed slide clamp sets require software 4.02.06 or higher.
35700BAX: 5.02.06, 6.02.06, 6.02.11Named in a noticeListed in the 2014 correction for pumps that one service technician may not have serviced correctly (Z-2738-2014). That notice covered specific serial numbers only.
V6: 6.02.xx and 6.05.xxV6, Master Drug Library Version 6Our hosted Operator’s Manual is for v6.05 with MDL Editor v6.2.4. Versions we commonly see on V6 pumps include 6.02.07, 6.05.13 and 6.05.14.
V8: 8.00.02Corrected 2017A software anomaly could cause a Sharp Watchdog Timeout error (Z-1693-2017).
V8: 8.01.01ReversionAn increase in false upstream occlusion alarms followed upgrades to this version; Baxter is reverting affected pumps at no charge (Z-2103-2023).
Master Drug Library 8.0Corrected 2015Loading and bolus dose times could display differently on the pump than set in the library during a four-site limited launch (Z-1451-2015).
V8 wireless battery module radio software 22D28, 22D29, 22D30UpgradeMay fail to auto-document infusion status to the EMR; Baxter is scheduling a software upgrade (Z-2581-2024).

FDA records show some 35700BAX pumps on pre-6 software such as 5.02.06, and the manual mentions 4.02.06; the “begins with 6 or 8” check in Baxter’s 2025 notice is for V6 and V8 software. Baxter has not published a current-version list that we could verify, and the 2020 correction for a programming error (Z-0712-2021 to Z-0714-2021) does not name versions. Confirm the current release for your fleet with Baxter, and contact us if you want help reading what your pumps show.

How to Check Your Sigma Spectrum Software Version

  1. Turn the pump on and note the software version on the start-up screen.
  2. On a V6, you can also open the Pump Information screen and press the sw info soft key, which lists the Sharp, PIC, CPLD, charger and network module versions too. On a V8, from the Select Care Area screen press the options menu soft key, then select User Options, View Information and Pump Information (press OK after each) to open the Pump Info screen.
  3. Confirm the version matches the platform: V6 software begins with 6 and V8 software begins with 8. V6 pumps typically carry product code 35700BAX or 35700ABB (FDA records also list variants such as 35700BAXR), and V8 pumps 35700BAX2. The software version is the deciding check.
  4. Record the version and serial number for each pump in your asset system so you can act quickly on future notices.

Sigma Spectrum Class I Recalls

Class I is FDA’s most serious recall classification. Eight Class I events have involved Sigma Spectrum V6 or V8 pumps:

InitiatedFDA recall numberPumpsIssueRemedy
September 2010Z-0146-2011Spectrum 35700, serials 706497–724065Possible bearing failure causing inaccurate flow, from back flow to free flow and over-infusion.SIGMA serviced or replaced affected pumps, which had to be returned.
July 2011Z-1490-2012Expansion to pumps made January 18, 2005 to November 1, 2010 (serials 700000–794213)Same bearing failure, linked to loss or degradation of bearing lubricant.Remove from neonatal use until B. Braun BC 1000 back check valves are used; affected pumps serviced.
February 2014Z-1484-2014V6 (35700BAX, 35700ABB)Repeated System Error 322 occurrences.Contact Baxter; its technicians determine whether adjustment or repair is needed.
June 2020Z-0120-2021, Z-0121-2021, Z-0122-2021V6 and V8, all serial numbers (also Spectrum IQ)Improper cleaning can cause residue or corrosion; on battery power alone, the pump may shut down without an alarm.Updated instructions for use. Inspect rear case pins and battery contacts and contact Baxter if residue or corrosion is found; keep backups for critical infusions; keep pumps plugged in when possible.
December 2021Z-0530-2022V8, all serial numbers (also Spectrum IQ)Reduced or non-delivery, sometimes without an alarm, from incorrect set setup or incomplete resolution of upstream occlusion alarms.Follow the manual’s setup and alarm steps. Do not press RUN/STOP before inspecting the tubing and resolving the occlusion.
June 2023Z-2103-2023V8 on software v8.01.01 (also Spectrum IQ v9.02.01)An increase in false upstream occlusion alarms after the software upgrade.Continue use per the manual; unload and reload the set if an alarm will not clear. Baxter schedules a software reversion at no charge.
February 2025Z-1253-2025V6, 66 listed serial numbers (also Spectrum IQ)Possible missing motor mounting screws from the servicing process.Isolate and stop using affected pumps; return them to Baxter for inspection and reservicing.
June 2025Z-2095-2025 (V6), Z-2096-2025 (V8)14 V6 and 28 V8 listed serial numbersThe pump may have the other platform’s software installed, which can cause flow inaccuracy and user confusion.Remove from service, check and record the software version, and contact Baxter to confirm it and schedule service.

The 2020 cleaning correction and the 2021 V8 upstream occlusion notice remain the most relevant to daily practice, because they applied to all serial numbers and depend on how pumps are cleaned and how alarms are handled. The 2025 notices apply only to listed serial numbers, but checking that every pump’s software matches its platform takes seconds.

Class II and III Recalls and Corrections

These Class II events involved V6 or V8 pumps, their software, their batteries or their accessories:

InitiatedFDA recall numberIssue
June 2013Z-0197-2014Reports of upstream occlusion and air-in-line alarms during infusions (V6). Follow the manual’s guidance on contributing factors and contact Baxter if a sensor is not functioning properly.
September 2013Z-0189-2014Sharp LCD screens (55074-2, revision B) on listed pumps did not meet electrostatic discharge standards.
September 2014Z-2738-2014Listed pumps on software 5.02.06, 6.02.06 and 6.02.11 may not have been serviced correctly by one service technician between May 5 and June 3, 2014.
March 2015Z-1451-2015Master Drug Library 8.0: loading and bolus dose times could round differently on the pump display.
September 2016Z-0670-2017, Z-0671-2017, Z-0672-2017Speakers in some V6 and V8 rear case assemblies, and in spare V6 rear cases, could fail prematurely; Baxter replaced the rear case assemblies at no charge.
January 2017Z-1693-2017V8.00.02 software could cause a Sharp Watchdog Timeout error, which is more likely when poor network maintenance floods the pump’s event history log.
October 2017Z-0971-2018Excessive pumping mechanism wear from a lack of lubrication on certain V6 pumps, which could cause over- or under-infusion. Customers were told to follow Preventive Maintenance document 41394 Rev. A for their whole fleet and return pumps that failed the flow-rate accuracy test.
April 2018Z-1663-2018 to Z-1667-2018Release testing on listed V6 pumps was improperly documented; air-in-line tests at 400 and 800 mL/hr and an upstream occlusion test at 800 mL/hr were omitted.
May 2018Z-2112-2018, Z-2113-2018V6 and V8 instructions said to lower the primary bag 20 inches below the secondary, but the supplied hangers are 12.5 inches; pumps can continue to be used with the supplied hanger.
June 2018Z-1248-2019Flow-rate testing on listed V6 pumps ran for less than the required minimum time before release.
November 2020Z-0712-2021 to Z-0714-2021A potential software error during programming on V6 and V8 pumps; Health Canada’s notice for V8 pumps says it may occur when programming a repeat of a completed secondary infusion. Baxter installs new software at the next service, or its local service representative contacts the facility.
January 2021Z-1150-2021Battery performance problems after prolonged storage, covering V6 and V8 pumps (35700ABB, 35700BAX, 35700BAX2) and standard and wireless battery modules (also Spectrum IQ). Baxter advised storing batteries out of the pump at 2 or 3 bars of charge between 14 and 95 °F, recharging every six months, running the Service Manual’s Battery Life Test at routine maintenance, and running pumps with recurring battery alerts on AC power while contacting Baxter.
June 2024Z-2530-2024One V8 pump released with improperly performed testing.
July 2024Z-2581-2024, Z-2582-2024Wireless battery modules A/B/G/N used with V8 (and Spectrum IQ) may fail to auto-document to the EMR; document manually until Baxter upgrades the software.
November 2024Z-0775-2025, Z-0776-2025One V6 and one V8 pump released after repair without full testing, including flow testing.
November 2025Z-1121-2026Listed V8 pumps released from service with defective, low-viscosity grease on the cam and motor gears, which could cause premature mechanism wear, over-infusion or free flow. Stop using affected pumps.

One Class III event also applies: in April 2012, SIGMA issued a customer safety alert for all versions of the Master Drug Library (MDL) Editor software, because it could produce a facility drug library that does not reflect common clinical practice (Z-2030-2012).

Many of these surface at the bedside as upstream occlusion alarms, system errors or battery problems. Our Sigma Spectrum common problems guide covers what we find when those pumps reach our lab.

What to Do Now: A Checklist for Biomed and HTM Teams

  1. Record the software version, product code and serial number of every Spectrum, and confirm V6 software begins with 6 and V8 with 8.
  2. Check your serial numbers against the listed-serial notices, especially Z-1253-2025, Z-2095-2025, Z-2096-2025 and Z-1121-2026.
  3. Inspect rear case pins and battery contacts for residue or corrosion at every PM, and confirm your cleaning agents and methods match Baxter’s current instructions.
  4. Make sure V8 users know not to press RUN/STOP until an upstream occlusion has been found and cleared.
  5. For V8 pumps on v8.01.01, confirm the reversion has been completed.
  6. Store spare batteries as Baxter advised in 2021: out of the pump, at 2 or 3 bars of charge, between 14 and 95 °F, recharged every six months.
  7. Run annual PM with flow-rate accuracy, air-in-line and occlusion checks on every pump.
  8. Check Baxter’s notices and FDA’s recall database for anything newer than this guide.

How IPR Medical Can Help

Version guidance

We can check your pumps’ software versions against the published notices. We update pump software as part of service; ask us about version upgrades. Recall remedies themselves come from Baxter.

Full evaluation in-house

We evaluate, repair and test Sigma Spectrum V6, V8 and Spectrum IQ pumps at our ISO 9001:2015-certified facility in Tustin, California, and verify flow-rate accuracy after every repair.

Battery contacts and sensors

We repair corroded charging pins, battery contacts, upstream pressure sensors, latches and mechanisms.

Free evaluation and shipping label

U.S. customers get a free inbound label and a free evaluation. Request a free label.

Need pumps instead? Shop recertified (refurbished) Sigma Spectrum V8 pumps or recertified Sigma Spectrum V6 pumps, or find manuals in our Sigma Spectrum manuals guide and Manuals & Documentation Library. For repairs, see our Baxter infusion pump repair page and our Sigma Spectrum repair guide.

Sigma Spectrum Recall FAQ

Is the Sigma Spectrum recalled?

Sigma Spectrum V6 and V8 pumps have been the subject of many FDA recalls and corrections since 2010, eight of them Class I. Most were corrections, such as updated instructions, software changes or service for listed serial numbers, rather than removal of every pump. The most recent Class I recall, from June 2025, covers specific serial numbers that may have the wrong software installed.

What was the Sigma Spectrum software mismatch recall?

In June 2025, Baxter recalled 14 listed V6 and 28 listed V8 pumps (Z-2095-2025 and Z-2096-2025) that may have the other platform’s software installed. Users were told to remove those pumps from service, check that V6 software begins with 6 and V8 software begins with 8, and contact Baxter.

What was the Sigma Spectrum battery corrosion recall?

In 2020, Baxter issued a Class I correction (Z-0120-2021 to Z-0122-2021) because improper cleaning can leave residue or corrosion on the rear case pins and battery contacts, which could cause a pump on battery power to shut down without an alarm. Baxter updated the instructions for use and asked facilities to inspect the pins and contacts on all pumps.

What should V8 users do about upstream occlusion alarms?

Baxter’s December 2021 Class I notice (Z-0530-2022) says not to press RUN/STOP before inspecting the tubing and resolving the occlusion, because the pump could appear to infuse while delivering less or nothing. In 2023, Baxter also began reverting V8 pumps on v8.01.01 because of false upstream occlusion alarms.

How do I check the software version on a Sigma Spectrum?

Turn the pump on and read the version on the start-up screen. On a V6, the Pump Information screen’s sw info soft key also shows it; on a V8, from the Select Care Area screen press the options menu soft key, then select User Options, View Information and Pump Information (press OK after each) to open the Pump Info screen.

Who makes the Sigma Spectrum now?

Baxter. The pump was developed by SIGMA International of Medina, New York. Baxter signed a distribution agreement and took a 40 percent stake in 2009, then bought the remaining 60 percent in April 2012.

Can IPR Medical check my pumps against the recalls?

Yes. We can check software versions against the published notices and evaluate pumps for the hardware issues they describe. Recall remedies such as software reversions and returns for reservicing are provided by Baxter, so for any pump on a recall list, contact Baxter as its notice directs; Baxter’s notice lists the affected serial numbers. Contact us about your fleet.