CADD Solis 2100

The CADD-Solis® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-Solis® pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia).

Features
  • Small and lightweight pump for convenience
  • Programmed intermittent bolus
  • Innovative therapies and task-based menus
  • Scalable platform
  • User-friendly interface
  • Graphical reports
  • Safety software

CADD Solis

So here you will be able to find the dimensions, weight, resolution, moisture protection, power and internal memory backup of CADD Solis Infusion Pump Repair. Download cadd solis pump cheat sheet and cadd solis vip pump video

  • Dimensions: 1.6″ W X 4″ H X 5″ L
  • Weight: 1.31 lbs. (21 oz), including 4 AA batteries
  • Resolution: Medication Cassette Reservoir or CADD Administration Set, 0.050 ml per pump stroke nominal
  • Moisture Protection: Splash-proof (IPX4) per IEC 60529
  • Power: 4 AA alkaline batteries; AC adapter; rechargeable battery pack
  • Internal Memory Backup: It charges whenever the pump is powered on and has a 10 month memory capacity once it has been charged for 250 hours at 68°F
  • System Operating Temperature: 36°F to 104°F (2°C to 40°C)
  • System Storage Temperature: -4°F to 140°F (-20°C to 60°C)
  • Atmospheric Pressure: 70 kPa (10.2 psi) to 106 kPa (15.4 psi)
  • High Pressure Alarm Threshold: 18 ± 9 psi
  • Air Detector Alarm:

Sensitivity:

    • Low – Single bubble greater than 400 uL
    • High – Single bubble greater than 150 uL

Accumulated Air – Greater than 1 ml air over 15 minutes (nominal)

  • Maximum Volume Infused Under Single-Fault Conditions: CADD® administration set: 0.15 ml
  • Delivery Rate during priming: Approx. 250 ml/hr
  • Alarm disabled during priming: Air-in-Line

Alarms/Alerts:

Battery depleted while delivering Reservoir volume empty
Battery removed while delivering Pressure sensor faulty
Battery unusable while delivering Downstream occlusion
Rechargeable battery end of life Upstream occlusion

 

Related Keyphrase:

  • CADD Solis Infusion Pump Repair
  • Download cadd solis pump cheat sheet
  • cadd solis vip pump video