Skip to content
IPR Medical
Kangaroo ePump Feeding Pump
Enteral Feeding PumpsManuals & DocumentationKangaroo ePump
Quick answer

You can download the Kangaroo ePump Operating Manual (Covidien, Manual No. SMF0211-079, REV 03/2012) and Cardinal Health’s two-page Quick Reference Guide for loading and rinsing below. We could not find a newer edition of the Operating Manual published on Cardinal Health’s website. The ePump is a rotary peristaltic enteral feed and flush pump that delivers 1 to 400 mL/hr in 1 mL steps with ±10% accuracy (or 0.5 mL/hr, whichever is larger), runs at least 15 hours on its NiMH battery at 100 mL/hr, and uses only Kangaroo ePump Feed-Only or Feed & Flush sets.

Whether you are training staff, checking an alarm at the bedside or setting up a biomed PM program, the right Kangaroo ePump document saves time. This guide collects the documents we host, explains which one you need, and pulls out the specifications and reference details people ask us about most.

Download the Kangaroo ePump Manuals

Operating Manual (REV 03/2012)

The 38-page Covidien Operating Manual for the Kangaroo ePump Enteral Feed and Flush Pump with Pole Clamp (Manual No. SMF0211-079): safety, setup, set loading and priming, continuous and intermittent feeding, flushing, the BioTech menu, performance evaluation, cleaning, battery replacement, error screens, specifications, part numbers and warranty.

Download PDF

Quick Reference Guide: Loading and Rinsing

A two-page Cardinal Health Canada guide (English and French, dated 11/12/2018) showing how to load and unload a set and how to rinse a set while it stays loaded in the pump.

Download PDF

These documents are provided for reference. They are the original manufacturer’s documents and are not maintained or updated by IPR Medical. For more manuals, browse our Manuals & Documentation Library.

Which Kangaroo ePump Document Do You Need?

The Operating Manual does not print a software version on its cover, so record the SW version from your pump’s BioTech screen (see below) and ask Cardinal Health whether a different edition applies to your pumps. Here is how the documents fit together:

DocumentWho uses itNotes
Operating Manual (SMF0211-079, REV 03/2012)Clinicians, educators, biomedThe full reference for programming, alarms, cleaning, performance testing and specifications. Published by Covidien before the Kangaroo line moved to Cardinal Health.
Quick Reference Guide (loading and rinsing)Clinicians and home caregiversSummary only; not a substitute for the full manual.
Performance test procedure and certificateTrained biomedical techniciansThe manual says the full test procedure and certificate are on the User Manual CD as a separate document. Not hosted here; the flow-rate accuracy check is summarized below.
Set instructions for useCliniciansEach Kangaroo ePump set has its own labeling; follow it along with the pump manual.

Checking recalls or end-of-life dates? Our Kangaroo ePump recalls and safety notices guide covers the FDA-listed set recalls and Cardinal Health’s discontinuation timeline.

What’s Inside the Operating Manual

  • Safety and warnings: approved AC adapter and sets only, the serial port warning, and the caution that premature infants may need higher accuracy than the pump provides.
  • Setup: attaching the AC adapter, connecting the battery, and mounting the pole clamp or cord clip.
  • Instructions for use: loading sets, auto-prime and hold-to-prime, continuous and intermittent (bolus) feeding, flushing, EZMODE, Keep Tube Open (KTO), Lock Screen, Hold, Flush Now and 72-hour history.
  • BioTech menu: serial number, software and flash versions, and the Lock Settings, EZMODE and Communication options.
  • Performance evaluation: the yearly test recommendation and the flow-rate accuracy check.
  • Cleaning, battery replacement and maintenance: approved cleaners, rotor cleaning, battery pack and door replacement.
  • Error screens and troubleshooting: the System Error number table and each alarm screen with its remedy.
  • Specifications, part numbers and warranty.

Key Kangaroo ePump Specifications

From the Operating Manual (REV 03/2012). Where the manual gives two figures for the same item, both are shown.

SpecificationKangaroo ePump
Pump typeVolumetric, rotary peristaltic
Model numbers382400 (U.S.), 482400 (international)
Feed rate1 to 400 mL/hr in 1 mL increments; BOLUS MAX delivers at 999 mL/hr
Feed volume to be delivered (VTBD)1 to 3000 mL in 1 mL increments (0 runs until the supply is exhausted)
Intermittent feeding1 to 3000 mL per bolus, 1 to 99 boluses, 1 to 24 hour intervals
Flushing10 to 500 mL per flush, every 1 to 24 hours; flushes at 1960 mL/hr
Accuracy±10% or 0.5 mL/hr, whichever is larger, with the bag 46 cm (18 in) above the pump at 22 °C ± 2 °C, using a new set with less than 24 hours of use
Occlusion pressure15 psi (103 kPa) nominal
BatteryReplaceable NiMH pack. Specifications: at least 15 hours at 100 mL/hr; Instructions for Use (General Startup): 15 hours at 125 mL/hr. Eight hours to fully recharge.
Power9 V DC, 1.5 A from the Kangaroo ePump AC adapter
HistoryUp to 72 hours of feed and flush totals, excluding the current hour
Alarm volumeAt least 70 dBA at 1 meter at maximum volume
Size and weight16.8 × 16.3 × 11.7 cm (6.6 × 6.4 × 4.6 in); 1.1 kg (2.4 lb), 1.2 kg (2.7 lb) with pole clamp
HousingABS/polycarbonate blend; set door polyester/polycarbonate blend
Electrical protectionClass II, internally powered; Type BF
Fluid protectionDrip-proof, IPX1 (IP40 also listed)
Operating conditions10 to 40 °C (50 to 104 °F), 75% RH non-condensing
Transport and storage0 to 50 °C (32 to 122 °F), 95% RH non-condensing

Pump Sets and Set Identification

The manual is direct: use only Kangaroo ePump sets, because the pump is not compatible with other pump sets. The pump reads the set type through MISTIC (Magnetic Intelligent Set Type Identification Connector), the black ring retainer, so the screen matches a Feed-Only or Feed & Flush set. Key points from the manual and Quick Reference Guide:

  • Valve into the left pocket with the finger tab aligned to the locating rib, tubing wrapped counterclockwise around the rotor without overstretching, retainer into the right pocket, then close the blue door. The status line should read SET LOADED.
  • Hang the bottom of the bag 18 in (46 cm) above the top of the pump for rated accuracy.
  • With a Feed & Flush set, prime the flush line first, then the feed line.
  • Do not reuse sets after 24 hours of use; the pump shows a blinking indicator once a set has run 24 hours.
  • To unload, power the pump off and wait 10 seconds before removing the set.
  • Never rotate the valve stem by hand.

Important: Cardinal Health’s June 2024 end-of-life letter said it anticipated that supply of ePump feeding sets and accessories would end June 30, 2025. See our recalls and safety notices guide for the full timeline.

Finding the Software Version: The BioTech Screen

  1. Press POWER to start the pump.
  2. Press and hold the top menu button while the start-up sequence runs (the kangaroo icon hops across the screen). You must still be holding it when the sequence ends.
  3. Read SN (serial number), SW (software version), FLASH (flash data version) and CERT INTV (power-ons since certification).
  4. Press DONE to exit. To return, power down and repeat.

The BioTech screen is also where biomed sets Lock Settings, EZMODE and Communication. Selecting CLEAR SETTINGS at power-up does not clear BioTech options such as EZMODE and Lock Settings.

Quick Reference from the Manual

TopicWhat the manual says
Preventive maintenanceRecommended service interval of at least once per year; run the performance evaluation yearly or whenever performance is suspect.
Flow-rate accuracy checkOn AC power with a new Feed-Only set and water, bag 18 in above the pump, run 15 minutes, then collect for 30 minutes. Expected: 33.7 to 41.3 mL at 75 mL/hr; 67.5 to 82.5 mL at 150 mL/hr.
BatteryConnect the battery before first use. Eight hours of charging fully recharges the pack; a replacement pack ships partially charged and should be charged 6 hours before use off AC power.
StorageFor storage longer than nine months, disconnect the battery; it may need several charge and discharge cycles to restore capacity.
BATTERY LOWAbout 15 minutes of battery life remain; plug in the AC adapter.
SYSTEM ERRORPower down and back on to try to clear it; if it persists, put a different pump into service and report the error number.
CleaningDisconnect AC power. Mild detergent for cleaning; 10% hypochlorite, isopropyl alcohol or ammonium chloride-based solutions for disinfection, on a damp (not wet) cloth. Do not immerse. Keep fluid out of the vents under the handle. Avoid Spray Nine, pHisoHex, Hibiclens, Vesta Syde and the other ingredients the manual lists.
SpillsAfter leakage into the pump, wetting of the AC adapter or a large formula spill, do not use the pump until it has been cleaned and serviced.
Serial portFor use only with the approved Communication Cable, part number 382499.

For what these alarms usually mean in practice, see our Kangaroo ePump common problems and repair guide.

User-Replaceable Parts

The manual says the main housing has no user-serviceable parts and opening it voids the warranty. It lists these customer-replaceable parts:

PartPart number
Main door (blue transparent door)F31877WT
Battery pack1041411 (manual); listed as 1054715 in Cardinal Health’s 2024 product list
Battery door (with screw)F31878WT
AC power adapter382491
Pole clamp382492
Electrical plugs (set of 4)382493

Service and Repair

IPR Medical evaluates, repairs, certifies and tests Kangaroo ePump pumps in-house at our ISO 9001:2015-certified facility in Tustin, California, including Covidien- and Kendall-labeled units. U.S. customers get a free inbound shipping label and a free evaluation, and nothing is charged until you approve the estimate. Learn more on our Kangaroo feeding pump repair page, or shop new and recertified (refurbished) Kangaroo ePump feeding pumps.

Kangaroo ePump Manual FAQ

Where can I download the Kangaroo ePump operating manual?

You can download the Covidien Kangaroo ePump Operating Manual (Manual No. SMF0211-079, REV 03/2012) and Cardinal Health’s Quick Reference Guide for loading and rinsing from this page.

How do I find the software version on a Kangaroo ePump?

Power the pump on and hold the top menu button until the start-up sequence ends. The BioTech screen shows the serial number, software version (SW), flash version (FLASH) and the number of power-ons since certification.

What is the flow rate range of the Kangaroo ePump?

1 to 400 mL/hr in 1 mL increments, with ±10% or 0.5 mL/hr accuracy, whichever is larger, under the manual’s test conditions. BOLUS MAX delivers boluses at 999 mL/hr, and flushes run at 1960 mL/hr.

How long does the Kangaroo ePump battery last?

The Operating Manual specifies at least 15 hours at 100 mL/hr from a new, fully charged NiMH pack (its General Startup section says 15 hours at 125 mL/hr). Eight hours of charging fully recharges it.

Can I use other feeding sets on a Kangaroo ePump?

No. The manual says to use only Kangaroo ePump sets and that the pump is not compatible with other pump sets. Cardinal Health’s 2024 end-of-life letter said it anticipated ePump set supply would end June 30, 2025.

How do I check Kangaroo ePump flow-rate accuracy?

On AC power, load a new Feed-Only set with water, hang the bag 18 in above the pump, run for 15 minutes, then collect for exactly 30 minutes. The volume should be half the hourly rate within ±10% or 0.5 mL, for example 33.7 to 41.3 mL at 75 mL/hr.

How often does a Kangaroo ePump need maintenance?

The manual recommends a service interval of at least once per year, plus performance testing any time the pump is suspected of not performing properly.

Need Help With a Kangaroo ePump?

Questions about a manual, an alarm or a repair? Our team can help, Monday through Friday, 7:00 AM to 4:30 PM Pacific.