Baxter Flo-Gard 6201
The Baxter Flo-Gard 6201 is a single-channel linear peristaltic volumetric infusion pump designed to deliver a wide variety of fluids — including blood and fat emulsions — across a broad range of infusion rates. A significant upgrade over the 6200, the 6201 adds incremental 0.1 mL/hr flow rate precision, a Programmed Delivery Profile (PDP) with up to 10 sequential infusion steps for ramped or manual programming, selectable air-in-line detection sensitivity, selectable downstream occlusion sensitivity, a Flow Check display, volume-time programming, a five-hour infusion data memory, and suitability for epidural administration.
It runs on 110/120V AC or an internal 12-volt sealed lead-acid battery with approximately 6 hours of runtime. Automatic piggybacking, a locking control panel, and an automatic safety clamp that occludes tubing on door open round out a pump that hospitals continue to rely on long after its discontinuation.
The Flo-Gard 6201 is an older, discontinued pump — but its dependability has kept it in active service across healthcare settings for decades. IPR Medical maintains a large inventory of Flo-Gard 6201 parts to keep your fleet running strong, from batteries and fuses to pump mechanisms and power cords.
Baxter Flo-Gard® 6201 Volumetric Infusion Pump
Single-channel linear peristaltic volumetric infusion pump for primary and secondary IV therapy including epidural administration. Delivers 0.1 mL increments at low rates and up to 1,999 mL/hr on primary, with 10-step Programmed Delivery Profile (PDP), selectable air-in-line and occlusion sensitivity, Flow Check display, RS-232 computer control, and 6-hour battery life. IPR Medical stocks a large inventory of Flo-Gard 6201 parts to keep your fleet running strong.
| Pump Type | Linear peristaltic volumetric infusion pump |
| Flow Rate — Primary | 1.0–99.9 mL/hr in 0.1 mL increments; or 1–1,999 mL/hr in 1 mL increments |
| Flow Rate — Secondary | 1.0–99.9 mL/hr in 0.1 mL increments; or 1–999 mL/hr in 1 mL increments |
| VTBI Range | 1.0–99.9 mL in 0.1 mL increments; or 1–9,999 mL in 1 mL increments (primary and secondary) |
| KVO Rate | 5 mL/hr or programmed rate, whichever is less |
| Programmed Delivery Profile (PDP) | Up to 10 sequential infusion steps; manual or ramp calculation |
| Ramp PDP — Total VTBI | 100–9,999 mL (run time ≤16 hrs); 200–9,999 mL (run time >16 hrs) |
| Ramp PDP — Delay Time | 00:00 – 08:00 hours in minute increments |
| Ramp PDP — Run Time | 01:00 – 24:00 hours in minute increments |
| Ramp PDP — Ramp Up | 00:00 to 45% of run time, in minute increments |
| Ramp PDP — Ramp Down | 00:00 to run time minus ramp up, in minute increments |
| Automatic Piggybacking | Yes — secondary program automatically switches to primary on completion |
| Volume-Time Programming | Yes — automatically calculates flow rate from volume and time entries |
| Air-in-Line Detection | Ultrasonic; NORM setting: ~75 µL or larger; MIN setting: ~50 µL or larger (selectable) |
| Occlusion Detection | Upstream and downstream; downstream sensitivity selectable via configuration |
| Free-Flow Protection | Automatic safety clamp occludes tubing when pump door is opened |
| Flow Check Display | Yes — displays real-time resistance to flow from norm to occlusion threshold |
| Panel Lockout | Yes — locking control panel prevents patient tampering |
| Memory | 5-hour retention of infusion data after power-off |
| Computer Interface | RS-232-C; remote control and monitoring capability |
| Nurse Call | Available via communications port (9-pin D connector) |
| Administration Set | Baxter designated “s” suffix standard sets; luer lock recommended for central lines |
| Epidural Use | Suitable for epidural administration |
| Power — AC | 110/120V, 60 Hz |
| Power — DC (Battery) | 12V, 2.0 Ah sealed lead-acid rechargeable |
| Battery Life | Approx. 6 hours at rates from 1 to 1,400 mL/hr |
| Battery Recharge Time | 8 hours for >80% recharge; 16–72 hours for full charge (depending on battery condition) |
| Power Cord | 2.9 m (9.5 ft), hospital-grade plug, replaceable |
| Fuse | 0.5A, 250V, SB, 5.2 mm × 20 mm; user-replaceable |
| Leakage Current | Less than 50 microamps (per UL 544) |
| Dimensions (W × D × H) | 20 cm × 13 cm × 29 cm (7.9 in × 5.1 in × 11.4 in) |
| Weight | Approx. 5.3 kg (11.6 lbs) |
IPR Medical maintains a large inventory of Flo-Gard 6201 parts — including batteries, fuses, power cords, pump mechanisms, and more. Contact us to keep your fleet running strong.
Trademark Notice: Baxter, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Baxter.
The Flo-Gard 6200 — taken further
The Baxter Flo-Gard 6201 builds on the proven foundation of the Flo-Gard platform with a meaningful set of clinical enhancements. Where the 6200 delivers in 1 mL/hr increments, the 6201 steps down to 0.1 mL/hr precision in the low-rate range — important for patients requiring careful titration. It adds the Programmed Delivery Profile for multi-step and ramped infusions, selectable air-in-line and occlusion sensitivity, volume-time programming, a Flow Check display, RS-232 computer interface, nurse call capability, five-hour data memory, and suitability for epidural administration. For facilities that relied on the Flo-Gard platform, the 6201 is the more capable version — and it remains in service across hospitals, long-term care facilities, and outpatient clinics for exactly that reason.
Programmed Delivery Profile — up to 10 sequential steps
The 6201’s Programmed Delivery Profile (PDP) allows clinicians to pre-program up to 10 sequential infusion steps, either manually or using the pump’s automatic ramp calculation. In ramp mode, the pump calculates flow rates across configurable delay, run, ramp-up, and ramp-down time segments — useful for tapering medications or managing complex fluid protocols. Volume-time programming lets the pump automatically calculate the flow rate when the total volume and desired infusion time are entered, reducing manual calculation and programming time at the bedside.
Configurable safety and detection
The 6201 features an ultrasonic air-in-line detector with selectable sensitivity — NORM mode alarms on bubbles approximately 75 µL or larger, while MIN mode detects bubbles as small as approximately 50 µL. Downstream occlusion sensitivity is also configurable through the pump’s settings. An automatic safety clamp occludes tubing any time the pump door is opened, preventing accidental free flow during set changes. The Flow Check display provides a real-time visual indicator of resistance to flow between the normal baseline and the next occlusion threshold — giving clinicians an early warning of developing downstream restrictions before a full occlusion alarm fires.
Computer interface, nurse call, and data memory
The 6201 includes an RS-232-C communications port that allows remote computer monitoring and control, as well as a nurse call output via the same 9-pin D connector. A five-hour infusion data memory retains programming after power-off — so a brief power interruption does not require full reprogramming when the pump is restarted. The locking front panel prevents patient tampering, and an automatic self-test routine runs on startup to verify proper function before each use.
IPR Medical keeps your Flo-Gard 6201 fleet running
The Flo-Gard 6201 is a discontinued product, but its reliability has kept it in active service long past its end-of-production date. IPR Medical maintains a large inventory of Flo-Gard 6201 replacement parts — batteries, fuses, power cords, pump mechanisms, and more — so your units stay in service rather than sitting on a shelf waiting for parts. Every pump we sell is fully inspected, tested, and patient-ready before it ships.
Need parts, repairs, or fleet pricing? IPR Medical stocks a large inventory of Baxter Flo-Gard 6201 components. Contact us for parts availability, repair turnaround, and volume pricing on units and parts.
Trademark Notice: Baxter, and all other brand names, model names, and part numbers shown are trademarks of their respective owners and are used on this page only to identify the original manufacturer’s equipment offered here, consistent with nominative fair use. This listing is offered by IPR Medical, an ISO 9001:2015-certified independent service organization, and is not sponsored by, affiliated with, or endorsed by Baxter.
| Weight | 14 lbs |
|---|---|
| Dimensions | 14 × 14 × 14 in |
| Product Condition | Refurbished / Patient Ready |